Dual-Channel Reference Thermometer Readout
Additel 282
OVERVIEW
Additel’s 282 Reference Thermometer Readout delivers the best possible accuracies and features in the palm
of your hand! With accuracy capabilities on par with laboratory grade thermometers, the ADT282 is capable
of handling even your most critical measurements. This ultra-high precision readout features dual analog
channels designed to facilitate comparison measurements and meet all of your temperature measurement
needs. The easy to use touchscreen makes navigating the well-designed menus a time saving and enjoyable
experience. The Lemo style smart connectors help to ensure that your probe calibration information is never
in question. The ADT282 Reference Thermometer Readout helps makes metrology simple and will quickly
become your new go-to when reliable temperature measurements are a must.
Accuracy to 0.006 °C at 0 °C
Dual Measurement Channels
RTD and TC Inputs
Smart Style Probe Connections
Large Smartphone Like Touchscreen
Differential Measurement Technology
Bluetooth & USB Communications
Built-in Sensor Library
Datalogging
IP67 Rated
Rugged Handheld Construction
Rechargeable Lithium Battery
Main Features
1mk temperature resolution, 0.1mΩ/0.1uV electrical measurement resolution
Reliable temperature metrology requires a stable repeatable measurement
device. The ADT282 supports stability and uniformity testing of liquid
temperature baths, thermocouple calibration furnaces, and dry well
calibrators. The highly accurate dual channels of the ADT282 support
deviation and uniformity studies. A very capable standard is required for
the measurement resolution of the thermometer readout in these situations.
The ADT282's superior measurement performance and dual-channel
conguration easily meet these measurement needs.
*Read up to two channels simultaneously
Temperature Calibration Equipment
02
Corporate Headquarters
2900 Saturn Street #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Email: sales@additel.com
Rev # 20220613
Main Features
Dual Channels
The model ADT282 includes dual inputs which provide support for a
multitude of resistive type sensors (RTD’s) as well as thermocouples (TC’s).
Both channels can be displayed simultaneously which allows for comparison
measurements and a host of other statistical analysis capabilities. And the
Additel 282 allows for easy differential measurement selection for T1-T2.
Reference Measurement Technology
Smart Style Probe Connections
One Touch Control Center
Datalogging
BlueTooth
Sensor Library
Additel’s 282 Reference Thermometer Readout utilizes a ratio measurement
technology which provides an unmatched performance in stability and drift. In
order to ensure a very small temperature drift coefcient and reliable long-term
stability, the ADT282 uses current reversal techniques to cancel EMF effects
and a ratio technology to cancel the A/D converter offset. This highly advanced
technology has not been available in a handheld device until now!
In order to facilitate quick and reliable probe connections, the ADT282 has
been configured with smart connection ports for probes. Both channels
atop the reference readout utilize 6 pin Lemo style smart connectors for
RTD probes and mini-TC ports for thermocouple probes. The thermocouple
connection points utilize an imbedded temperature sensor which allows for
both internal and external cold junction compensation.With the smart probe
connectors, the ADT282 utilizes a user-selectable probe lock feature to pair
the probe with the channel it was calibrated with in a system calibration.
In order to improve the user's experience and speed of use, we have designed
a single touch menu option that navigates users to a control center panel. The
functions from the control panel include: Date, Battery status, Screen lock,
Bluetooth on/off, Speaker on/off, Snapshot, Smart diagnosis center button.
The ADT282 comes with standard Bluetooth communications capabilities
and is supported by Additel’s Mobile Link App. This very useful feature
will change the way you work as it provides a remote view of the ADT282
display at a distance, up to 20 meters on your personal mobile device.
Temperature sensors and instruments used in the field often require regular
calibration. In many cases, the disassembly of equipment can impact
productivity. Fixed sensors can be tested in process utilizing the ADT282
datalogging capabilities. In order to accurately monitor temperature changes, this
process may take several minutes or even hours to complete. ADT282 has built-
in powerful data logging function and supports multi-parameters data recording,
trend curve display, partial curve observation, statistical result viewing, data
storage capacity up to 8G to help with these applications.
The ADT282 has an extensive built-in temperature sensor library, including
ITS-90, CVD, Standard TC,13 types of industrial RTDs and 15 types of
industrial thermocouples, and also supports sensor customization. The user
can also edit the probe coefcients according to the ITS-90, CVD formulas
and the R0 parameter of the industrial RTDs. The extensive probe library
capabilities also support coefcient input methods for standard thermocouple
types.
Temperature Calibration Equipment
03
Corporate Headquarters
2900 Saturn Street #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Email: sales@additel.com
Rev # 20220613
SPECIFICATIONS
General Specifications
Technical Specifications
Display 5.0 inch 480 x 800 TFT LCD capacitive screen
Size 16.97" x 4.13" x 2.04" (177 mm x 105 mm x 52 mm)
Weight 1.5 lbs. (0.65 Kg)
Power Supply
6600mAh, 23.8Wh lithium battery, charging time 4~6 hours, battery pack can be charged independent. Battery life
typically 16 hours
Environment
Specication guaranteed temperature range: (10~30) °C
Working Temperature: (-10~50) °C
Storage temperature: (-20~70) °C
Humidity: 0% ~ 95% RH, non-condensing
Warm-Up Time 10 minutes
Ports Protection Voltage 50V max
CE Certicate TUV IEC61326
,
IEC61010
Rohs Compliance Rohs II Directive 2011/65/EU
EN50581:2012
IP Protection Level IP67, 1 meter drop test
Communication Isolate USB-TYPEC (slave), Bluetooth BLE
Input Channels CH1, CH2 analog channel, 6 pins smart lemo ports for RTD probe
MINI-TC ports for TC probe
Measurement Display Single channel, dual channel, differential (e.g T1-T2)
Measuring Rates
CH1, CH2 analog channels alternately and cyclically measure
RTD measuring rate: 1.6S/single channel, 1.6S/dual channel
TC measuring rate: 0.8S/single channel, 0.8S/dual channel
Measurement Units °C, °F, K
Statistics Max, Min, Avg
Warranty 1 Year
Specification
PRT Measurement
RTD Types
ITS-90, CVD, Ohms, Pt100 (385), Pt10 (385), Pt25 (385), Pt50 (385), Pt100 (3916),
Pt100 (3926), Pt100 (391), Cu100 (428), Cu50 (428), Cu10 (427), Ni100 (617),
Ni100 (618), Ni120 (672), and custom RTD
Resistance Accuracy 0~400Ω: ±0.5mΩ@(0~20Ω),±25ppm@(20~400Ω
Measurement range -200°C ~ 850°C
Resolution ±0.1mΩ or 0.001°C
Connection Type 4-wire smart connection
Excitation Current 1 mA - alternating constant current
Temperature Coefcient ±2ppm FS/°C (-10°C~10°C and 30°C~50°C)
TC Measurement
TC Types mV, S, R, B, K, N, E, J, T, C, D, G, L, U, LR, A, 10uV/°C, 1mV/°C, Standard TC
Electrical Measurement -10~75mV: 50ppm RDG+2uV
TC measurement range -270°C ~ 1800°C
Resolution ±0.1uV or 0.001°C
Connection Type Mini-TC
CJC compensation methods Internal, external or manual entry
Temperature Coefcient ±5ppm FS/°C (-10°C~10°C and 30°C~50°C)
Internal CJC Specication ±0.15°C (-10°C~50°C)
Measurement Specifications
04
Corporate Headquarters
2900 Saturn Street #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Temperature Calibration Equipment
Phone: 714-998-6899 Email: sales@additel.com
Rev #20220613
Thermocouple Measurement (Environment Temperature: 20±10
o
C)
Type Temperature Range (
o
C)
Accuracy (
o
C)
External CJC compensation
(1 year)
Accuracy (
o
C)
Internal CJC compensation
(1 year)
S -50 to 1768
-50~0 0.51 0.53
0~100 0.37 0.40
100~1768 0.28 0.32
R -50 to 1768
-50~0 0.54 0.56
0~200 0.38 0.41
200~1768 0.25 0.29
B 0 to 1820
200~300 1.01 1.02
300~500 0.66 0.68
500~800 0.41 0.44
800~1820 0.28 0.32
K -270 to 1372
-250 to -200 0.48 0.50
-200 to -100 0.15 0.21
-100 to 600 0.08 0.17
600 to 1372 0.14 0.21
N -270 to 1300
-250 to -200 0.76 0.77
-200 to -100 0.22 0.27
-100 to 1300 0.12 0.19
E
-270 to 1000
-250~-200 0.26 0.30
-200~-100 0.10 0.18
-100~700 0.06 0.16
700~1000 0.08 0.17
J -210~1200
-210~-100 0.13 0.20
-100~700 0.06 0.16
700~1200 0.10 0.18
T -270 to 400
-250~-100 0.36 0.39
-100~0 0.08 0.17
0~400 0.05 0.16
C
0 to 2315
0 to 1000 0.16 0.22
1000 to 1800 0.26 0.30
1800 to 2315 0.42 0.45
Accuracy Specifications
Accuracy (
°C
)
T, °C Rx,Ω
ADT282
readout only
(°C)
Readout with selected Probe Accuracy (
o
C)*
AM1760 AM1751 AM1730 AM1640 AM1660 AM1710 AM1612-2 AM1612-1
-200 18 0.005 0.013 0.021 0.021 0.053 0.053 n/a 0.072 n/a
-40 84 0.005 0.013 0.018 0.018 0.042 0.042 0.018 0.051 0.051
0 100 0.006 0.009 0.014 0.014 0.036 0.036 0.014 0.051 0.051
100 140 0.009 [1] [1] [1] [1] [1] 0.019 0.051 0.051
160 163 0.011 [1] [1] [1] [1] [1] 0.023 0.052 0.052
232 190 0.013 0.019 0.024 0.024 0.059 0.059 n/a n/a n/a
420 257 0.018 0.027 0.033 0.033 0.077 0.077 n/a n/a n/a
660 338 0.026 0.040 0.046 n/a n/a 0.109 n/a n/a n/a
Note: [1] - These are non-standard calibration points, some probes are not calibrated at 100°C and/or 160°C
.
*Includes readout accuracy, probe calibration, and probe drift.(K=2)
Thermocouple Measurement
05
Corporate Headquarters
2900 Saturn Street #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Temperature Calibration Equipment
Phone: 714-998-6899 Email: sales@additel.com
Rev #20220613
Accessories (Included)
Model Description QTY
9813-X
Power Adapter, external power adapter
for Additel 282 Thermometer Readout
1 pc
9052 USB Cable type A to type C 1 pc
9704 Chargeable Li-ion battery
1 pc
ISO 17025 accredited calibration
certicate
1 pc
Ordering Information
Accessories
Optional Accessories
Model Description
9070
Smart connector for reference PRT used with ADT875,
ADT878, and ADT282
9071
Connector Adapter from smart connector to 4-wire with
gold-plated spades for AM17XX PRTs
9072 Smart connector with clamps
9080
Cable kits (including TC plug, compensation cable,
S,R,K,J,T,E,N)
9918-SC
Soft carrying case, with space for instrument, test leads,
and accessories
9905
Carrying case for handheld calibrators and readouts with
space for two PRTs
SPECIFICATIONS
Thermocouple Measurement (Environment Temperature: 20±10
o
C)
Type Temperature Range (
o
C)
Accuracy (
o
C)
External CJC compensation
(1 year)
Accuracy (
o
C)
Internal CJC compensation
(1 year)
D
0~2315
0~100 0.21 0.26
100~1200 0.16 0.22
1200~2000 0.27 0.31
2000~2315 0.42 0.45
G 0 to 2315
50~100 0.60 0.62
100~200 0.38 0.41
200~400 0.24 0.28
400~1500 0.16 0.22
1500~2315 0.32 0.35
L -200 to 900
-200 to -100 0.07 0.17
-100 to 400 0.06 0.16
400 to 900 0.07 0.17
U -200 to 600
-200 to 0 0.14 0.21
0 to 600 0.05 0.16
LR -200~800
-200~0 0.09 0.17
0~800 0.06 0.16
A
0~2500
0~1200 0.20 0.25
1200~2000 0.33 0.36
2000~2500 0.48 0.50
Notes:
1. The index is based on the accuracy of the thermocouple electrical measurement, does not include the accuracy of the thermocouple itself
and the xed cold junction compensation at 0 °C.
2. Combined accuracy specications of probe and readout are calculated using the RSS method.
3. Additel provides standard S-typeTC probe with MINI-TC connector.
Model Number
ADT282
*See page #6 for ordering info regarding common probes
used with the ADT282.
06
Corporate Headquarters
2900 Saturn Street #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Temperature Calibration Equipment
Phone: 714-998-6899 Email: sales@additel.com
Rev #20220613
Specification AM1710 Series AM1730 Series AM1751 Series
Temperature Range -60°C to 160°C -200°C to 420°C -200°C to 670°C
Resistance at 0
Nominal 100Ω
Temperature Coefficient
0.003925 Ω / Ω / °C
Accuracy
±
0.025°C at -40°C
±
0.015°C at 0.01°C
±
0.025°C at 160°C
±
0.025°C at -196°C
±
0.015°C at 0.01°C
±
0.035°C at 420°C
±
0.025°C at -196°C
±
0.015°C at 0.01°C
±
0.035°C at 420°C
±
0.05°C at 661°C
Drift
±
0.01°C at TPW after 100 hours at
160°C
±
0.01°C at TPW after 100 hours at
420°C
±
0.01°C at TPW after 100 hours at
661°C
Short Term Stability
±
0.007°C
Thermal Shock
±
0.005°C after 10 times thermal cycles from minimum to maximum temperatures
Hysteresis <=0.005°C
Self-heating
50 mW/°C
Response Time 9 seconds for 63% response to step change in water moving at 3 feet per second
Measurement Current
0.5 mA or 1 mA
Sensor Length 32 mm
Sensor Location 5 mm from tip
Insulation Resistance >1000 MΩ at room temperature
Sheath Material
Stainless Steel Inconel
tm
Dimension
AM1710-12-ADT
0.25 in dia X 12 in (6.35 mm X 305 mm)
AM1730-12-ADT
0.25 in dia X 12 in (6.35 mm X 305 mm)
AM1751-12-ADT
0.25 in dia X 12 in (6.35 mm X 305 mm)
AM1710-BEND-ADT
0.25 in dia X 12 in (6.35 mm X 305 mm),
90° bend at 7.4 inch (190 mm) from probe
end
AM1730-BEND-ADT
0.25 in dia X 12 in (6.35 mm X 305 mm),
90° bend at 9.6 inch (245 mm) from probe
end
AM1751-BEND-ADT
0.25 in dia X 12 in (6.35 mm X 305 mm),
90° bend at 9.6 inch (245 mm) from probe
end
External Leads Teon
tm
–insulated copper wire, 4 leads, 2.5 meters
Handle Dimension 15 mm (OD) x 65 mm (L)
Handle Temperature Range
[1]
-50°C to 160°C -50°C to 180°C
Optional Calibration NIST traceable calibration and data available per request
[1] Handle temperature outside this range will cause damage to the probe.
* PRT Information from www.accumac.com
AM17XX-12-ADT AM17XX-BEND-ADT
Secondary PRT Information
AM1710
Secondary PRT Ordering Information
ADT
PRT Exterior:
12 - 12 inch straight
BEND - 90
° bend
12
Secondary PRT Model:
AM1710
AM1730
AM1751