Pressure Calibration Equipment
01
Corporate Headquarters
2900 Saturn St #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Email: sales@additel.com
Rev # 20230414
Automated Hydraulic Pressure Calibrator
Additel 762W
OVERVIEW
The Additel 762W Automated Pressure Calibrator provides a revolutionary turnkey
solution for automated pressure calibration work up to 10,000 PSI using deionized
water as the working media.
Designed for use in both the eld and the laboratory, the portability and accuracy of this
state-of-the-art product will quickly become the favorite go-to calibrator for lab personnel
and eld technicians alike.
With fully automated support for calibration of pressure transmitters, switches, dial
and digital gauges and sensors, including HART/PROFIBUS devices in conjunction
with a fully integrated task feature, data collection and Wi-Fi connectivity, we had our
customer’s needs in mind when designing our most capable pressure calibrator to date.
Automated pressure generation and control to 10,000 psi (700 Bar)
Accuracy to 0.01%FS
Dual-range manual or auto select
Designed for use with water
Automated fluid management
Control stability 0.005%FS
Portable - designed for use in the field and in the lab
Supports two external pressure modules
Wi-Fi, LAN, Bluetooth, USB and Ethernet communication
Full HART field communicator
HART and PROFIBUS communication
Data logging and task management
Patented electric pump technology
Corporate Headquarters
2900 Saturn St #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Email: sales@additel.com
Rev # 20230414
02
Pressure Calibration Equipment
Dual-Range Accuracy to 0.01% FS
The ADT762W includes the unique ability to automatically switch between
different internal calibrations depending on the current control pressure of the
ADT762W. Additel provides calibrations unique to each ADT762W for ranges
of 0-3,000 PSI (200 Bar) and 0-10,000 PSI (700 Bar). As the calibrator is
pressurized, it will automatically select the control and measurement specication
based on the specic pressure range. Pressure calibration range selection can
be set to “auto” mode so the calibration range is automatically selected by the
ADT762W based on the set point pressure, or the calibration range can be
manually selected.
Dual-Mode HART Communication
HART pressure transmitters can be directly maintained and calibrated without any other equipment or tools. The
ADT762W provides an automatic HART calibration mode as well as a manual mode. This dual-mode HART
communication function not only provides an efcient and convenient interoperability mode for DUT, but also supports
access to a fully HART capable calibrator.
Documented Task Feature
The powerful documented calibration task application allows users to quickly
create and execute tasks without the need for a PC or tablet. The ADT762W
automatically analyzes errors, generates test reports, while storing results locally.
The Additel 762W can support up to 1000 documented tasks which can be stored
and recalled at any time to help save time and money.
Built-in Auto-Purge Application
Purging hydraulic calibration systems can be challenging and time consuming.
The ADT762W has been specifically designed for use with water. It comes
standard with a specially designed manifold to help provide a constant positive
pressure environment to mitigate concerns with contamination when using
water as a medium. The integrated auto-purge system saves time, money and
frustration by completely automating the removal of air from the system. With the
push of a button, the ADT762W quickly manages the system purging. This helps
to free up time for technicians to attend to other needs.
Dual-Port Manifold with Automatic Fluid Extraction
When using hydraulic systems, we know that oils are a better lubricant and
sustain the life of seals and components than water. But we recognize that for
some applications, using oil is not an option. The ADT762W is designed to only
be used with deionized water. The ADT762W has a built-in fluid management
system to reduce standing water in the system which eases future maintenance
requirements. The ADT762W should be used with the provided manifold which
has a communication connection directly to the unit to actuate a vent valve built
into the manifold. This configuration is designed to prevent contaminated fluid
from returning to the base unit and manifold as it is deposited into a special waste
tank on the bottom of the ADT107 manifold. This function helps ensure the longest
life of the unit and reduced maintenance costs.
Pressure Calibration Equipment
03
Corporate Headquarters
2900 Saturn St #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Email: sales@additel.com
Rev # 20230414
Pressure Specifications
Electrical Specifications
Supported
[1] Control Stability is based on the range selection or external module
Model
Specification
Range Resolution Accuracy Note
mA Measure
-25 to 25 mA 0.1 μA ±
(
0.008%RD + 1.0 μA
)
Impedance <10 Ω
-50 to 50 mA 0.1 μA ±
(
0.008%RD + 2.0 μA
)
V Measure -300 to 300 mV 1 μV ±
(
0.008%RD + 6 μV
)
Impedance <1 GΩ
V Measure
(Auto-ranging)
-5 to 5 V 20 μV ±
(
0.008%RD + 100 μV
)
Impedance >1 MΩ
-12 to 12 V 100 μV ±
(
0.008%RD + 320 μV
)
-30 to 30 V 100 μV ±
(
0.008%RD + 600 μV
)
Loop Power Source 24 V N/A ±1 V 50 mA (Max Loading)
mA Source
0 to 2.5 mA or
2.5 to 25 mA
0-2.5 mA: 0.05μA
0-25 mA:0.5μA
0-2.5 mA: 0.008%RD + 0.2 μA
0-25 mA: 0.008%RD + 1.0 μA
20 mA @ 1 KΩ
Power Source 16 to 30 V 1 V ±1 V 70 mA (Max Loading)
V Source 0 to 16 V 0.25 mV 0.008%RD + 500 μV
Pressure Switch
Mechanical Switch, Live
Mechanical Switch, NPN Switch,
PNP Switch
N/A N/A
Response time<10 ms.
If the switch is live, voltage range
will be (3-30) V
Temperature Compensation 18 °C to 28°C
Temperature Coefficient Outside of 18 °C to 28 °C: <± 0.0005%RD + 0.00005%FS/°C
Misuse Protection Up to 30 V on any two sockets
Pressure Switch Test
HART / PROFIBUS PA
Model
Specification
ADT762W Automated Hydraulic Pressure Calibrator
Pressure Range 15~10,000 psi (1-700 Bar)
Range Selection Manual 3K psi, Manual 10K psi or Auto-range
Accuracy
0~10,000 psi, 0.01%FS 0~10,000 psi, 0.02%FS
0~3,000 psi, 0.01%FS 0~3,000 psi, 0.02%FS
Resolution 7 digits 6 digits
Measurement Units
Pa, hPa, kPa, mPa, bar, mbar, psi, mmHg@0°C, cmHg@0°C, mHg@0°C, inHg@0°C, inH2O@4°C,
mmH2O@4°C, cmH2O@4°C, mH2O@4°C, mmH2O@20°C, cmH2O@20°C, mH2O@20°C,
inH2O@20°C, inH2O@68°F, kgf/cm2, mtorr, torr, lb/ft2, tsi, custom
Maximum External Load Capacity Max: 80 ml@700 Bar, 50 ml recommend
Reservoir Max: 350 ml
,
built-in lter
Control Stability
[1]
0.005%FS from 100 to 10,000 psi
Stability Duration
5 min
Pressure Module Built-in one module with dual range
External Control Pressure Module See the following "External Control Pressure Module Specication and Compatibility" table
External Measurement Pressure Module All ADT161 pressure modules
Corporate Headquarters
2900 Saturn St #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Email: sales@additel.com
Rev # 20230414
04
Pressure Calibration Equipment
Optional Accessories
Model number Description Picture
ADT161
Pressure modules
see pg. 4
ORDERING INFORMATION
Accuracy:
01 = 0.01% of full span
02 = 0.02% of full span
Pressure port type:
N - 1/4 NPT female
N2 - 1/2 NPT female
B - 1/4 BSP female
B2 - 1/2 BSP female
M - M20X1.5 female
ADT762W
01
Model Number
GP10K
N
Accessories (included)
Model number
Quantity Picture
9022
Test Leads
2 sets
(4 pcs)
USB Cable 1 pc
9725
Rechargeable Li-ion battery
1 pc
9818
110V/220V external
Power adapter
1 pc
Funnel 1 pc
ADT100-762Hose
High pressure hose
1 pc
ADT100-762-SSTUBE
Stainless steel high pressure hose
1 pc
ADT107-X-KIT
(dual ports manifold, and zero return
communication cable)
1pc
Small accessory case
(For cables and adaptors)
1 pc
Transportation cap 1 pc
9907-762
Carrying case
1 pc
O rings for liquid storage tank
3.5*1.2-NBR70
5 pcs
O rings for right angle connector
5*1.5-NBR70
5 pcs
9060
Pressure module connection cable
1 pc
ISO17025 accredited
calibration certificate
1 pc
Pressure Calibration Equipment
05
Corporate Headquarters
2900 Saturn St #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Email: sales@additel.com
Rev # 20230414
External Control Pressure Module Specification and Compatibility
General Specifications
Specification Description
User Interface Color touch screen and keypad operation
Display 7” TFT touch screen 800 x 480 color
Enclosure IP Rating IP31
Power Dedicated lithium battery or power adapter
Battery Rechargeable Li-Ion battery, typically 12 hours of operation, less than 5 hours recharge.
Weight 28 lbs (12.8 kg) without media
Media Deionized water
Size 11.81 x 8.66 x 7.56 in (300 x 220 x 192 mm)
Communications USB, LAN, Bluetooth, Wi-Fi and Ethernet
HART Communication
Read, congure and calibrate HART devices - DD les updated periodically
Data Storage
8 GB
Data Logging Up to 1,000,000 readings (data and time stamped)
Task Documentation Up to 1,000 tasks can be stored with data
Automation Functions Switch test, auto step, leak test
User Interface Localization English, German, French, Italian, Spanish, Portuguese, Chinese, Japanese, and Russian
Pump Life > 1,000,000 cycles
Environmental Specifications
Operating temperature: 32 °F to 122 °F( 0 °C to 50 °C)
Storage temperature: -20 °C to 60 °C ( -4 °F to 120 °F)
Humidity: <90%, non-condensing
Certification ISO 17025 accredited certicate of calibration with NIST-traceable data
Compliance CE
Software Compatibility ACal, Additel Land and Additel Link for access via mobile application
Warranty 1 year
Specification
Model
Pressure Range
Accuracy
Pressure Type Media
(psi) (bar)
ADT161-01-GPXX
for 0.01%FS
1,000 70 0.01% FS Gauge G,L
1,500 100 0.01% FS Gauge G,L
2,000 140 0.01% FS Gauge G,L
3,000 200 0.01% FS Gauge G,L
5,000 350 0.01% FS Gauge G,L
10,000 700 0.01% FS Gauge G,L
ADT161-02-GPXX
for 0.02%FS
1,000 70 0.02% FS Gauge G,L
1,500 100 0.02% FS Gauge G,L
2,000 140 0.02% FS Gauge G,L
3,000 200 0.02% FS Gauge G,L
5,000 350
0.02% FS
Gauge G,L
10,000 700 0.02% FS Gauge G,L