Pressure / Process Calibration Equipment
01
Corporate Headquarters
2900 Saturn St #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Fax: 714-998-6999 Email: sales@additel.com
Rev # 20210616
Multifunction Documenting Process
Calibrators
Additel 221A, 222A & 223A
Sourcing, simulating and measuring pressure, temperature
and electrical signals
Smartphone-like menu and interface for simple operation
HART Communication capability (223A)
The internal cold junction compensation sensor can be re-
calibrated at the ice point
Ultra-compact, 3.9" x 7.6" x 2.0", and 1.6 lb (0.7 kg)
OVERVIEW
This series of highly integrated multifunction calibrators feature several patented technologies. These are an ultra-
compact, rugged, and easy to use hand-held device for sourcing, simulating and measuring pressure, temperature, and
electrical signals. Their smartphone-like menu and interface make the operation simple. Automation and documentation
capabilities make the these calibrators a turnkey solution.
Additel 221A
The Additel 221A is very unique to the multifunction temperature calibrator market. Not only does it provide you with the
ability to source, simulate and measure temperature and electrical functions but it also incorporates full documenting
capability and many other solutions other competitors do not provide. As a standalone device, you can create tasks, run
tests and store the results. With the use of Additel/Land Software or Additel/Cal software, all saved tasks and data can
be downloaded and managed. With it's unique internal cold junction compensation sensor, cold junction compensation is
very simple and easy.
Additel 222A
The Additel 222A Multifunction Process Calibrator takes all the functionality of the 221A and adds the ability to measure
and source pressure using the ADT160A Intelligent Pressure Modules. With calibration characteristics programmed
directly into the ADT160A pressure modules, you can simply connect them to the 222A and it will automatically recognize
and display the module as the source or measure pressure. For ultimate convenience, the 222A allows for modules to be
"hot swapped" with a simple connection at the top of the module.
Additel 223A
The 223A has all the capability of the 221A and 222A and HART Communication in one small, easy-to-use package.
The 223A has a full HART library that allows for the reading of HART-smart devices and also the capability to write to
devices. Combined with full task automation and documentation, the 223A is an ideal tool to accomplish many of your
important calibration tasks.
Sourcing, simulating and measuring temperature and electrical
signals
Sources and measures mV, mA, ohms, RTDs,
thermocouples, frequency, and pulses
Simulates and measures 13 thermocouples and 11
RTDs to calibrate transmitters
Measures and sources pressure using Additel 160A
series Intelligent Digital Pressure Modules from -15 psi
to 10,000 psi (-1 bar to 700 bar)
24V loop power supply
Simultaneous dual reading capability
Automatic switch test
Supports square root transmitter
Pulse frequency output for the calibration of ow totalizer
Easy to use
Smartphone-like menu and interface make the operation
simpler and easier
Ultra-compact, size 3.9" x 7.6" x 2.0" (100mm x 192mm
x 52mm), and weight 1.6 lb (0.7 kg)
One handed operation
Calibrated cold junction compensation (Patented)
Cold junction equalizing block in the calibrator
A calibrated PRT element with flexible leads is
installed
in the equalizing block for thermocouple cold junction
compensation
This PRT element can be pulled out from the calibrator
and re-calibrated and corrected at the ice point
Documenting and automated procedure capability
Manage the information of the device under test.
Set up automated calibration procedures, and 223A
performs the test, calculates the errors, displays and/
or stores the results in the memory, and highlights the
out-of-tolerance points.
As-found and As-left functions allow recording and
documenting results for quality control.
Download tasks and upload the results.
Snapshots allow you to capture and save data.
ISO 17025 accredited calibration with data
FEATURES
Pressure / Process Calibration Equipment
02
Corporate Headquarters
2900 Saturn St #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Fax: 714-998-6999 Email: sales@additel.com
Rev # 20210616
SPECIFICATIONS
Electrical Specifications
General Specifications
Measurement Accuracy
Range Resolution Accuracy
Voltage DC
±75.0000 mV
0.1μV 0.01%RD + 3.75 μV
±30.0000 V 0.1 mV 0.01%RD + 1.5 mV
Current DC
± 30.0000 mA 0.1μA
0.01%RD + 1.5 μA
Resistance
Two-wire 0 to 400.000Ω 1 mΩ 0.02%RD + 0.02Ω
Three-wire 0 to 400.000Ω 1 mΩ 0.02%RD + 0.02Ω
Four-wire 0 to 400.000Ω
1 mΩ 0.01%RD + 0.02Ω
Two-wire 0 to 4000.00Ω 10 mΩ 0.02%RD + 0.2Ω
Three-wire 0 to 4000.00Ω 10 mΩ 0.02%RD + 0.2Ω
Four-wire 0 to 4000.00Ω
10 mΩ 0.01%RD + 0.2Ω
Frequency 1 to 50000.0 Hz 0.1 Hz 0.005%RD + 1 Hz
Pulse
0 to 999999 1
N/A
Limit Switch
For the contact with potential, the voltage within the
range 3 V to 24 V.
Temperature Specification
Thermocouple Measurement and Source Accuracy
Measure
and Simulate
Standard Temperature Range (
o
C)
Accuracy (
o
C)
Measure Source
S IEC 584
-50 to
1768
-50 to400 1.0 1.1
400 to 1000 0.6 0.6
1000 to1768 0.7 0.8
R IEC 584
-50 to
1768
-50 to 200
1.4 1.4
200 to 500 0.6 0.6
500 to 1768 0.6 0.7
B IEC 584 0 to 1820
50 to 450 3.8 3.8
450 to 800 0.9 0.9
800 to 1820
0.6 0.7
K IEC 584
-270 to
1372
-250 to -200 1.0 1.1
-200 to -100 0.4 0.5
-100 to 600 0.3 0.3
600 to 1372 0.4
0.5
N IEC 584
-270 to
1300
-250 to -200 1.5 1.6
-200 to -100 0.5 0.6
-100 to 1300 0.4 0.5
E IEC 584
-270 to
1000
-250 to -200 0.6
0.7
-200 to -100
0.3
0.3
-100 to 0 0.2 0.2
0 to 700 0.2 0.3
700 to 1000 0.2 0.4
J IEC 584
-270 to
1200
-210 to -100 0.3 0.3
-100 to 1200 0.3
0.4
T
IEC 584
-270 to
400
-250 to -200 0.8 0.9
-200 to 0 0.4
0.4
0 to 400 0.2 0.2
C
ASTM
E988
0 to 2315
0 to 1000 0.5
0.5
1000 to 1800 0.7 0.9
1800 to 2315 1.0 1.4
D
ASTM
E988
0 to 2320
0 to 100 0.5 0.5
100 to 1100 0.4 0.5
1100 to 2000 0.6 0.9
2000 to 2320 0.9 1.3
G
ASTM
E1751
0 to 2315
0 to 200 2.4
2.4
200 to 400 0.5 0.5
400 to 1400 0.4 0.5
1400 to 2315 0.7 1.0
L
DIN43710
-200 to
900
-200 to -100 0.2 0.3
-100 to 400 0.2
0.2
400 to 900 0.2 0.3
U
DIN43710
-200 to
600
-200 to 0 0.4 0.4
0 to 600 0.2 0.3
Source Accuracy
Range Resolution
Accuracy
Voltage DC
-10.000 to 75.000 mV 1 μV 0.02%RD + 4.25 μV
0 to 12.0000 V 0.1 mV 0.02%RD + 0.6 mV
Current DC 0 to 22.000 mA 1 μA 0.02%RD + 1.1 μA
Resistance
1 to 400.00 Ω 10 mΩ 0.02%RD + 0.02Ω
1 to 4000.0 Ω 100 mΩ 0.03%RD + 0.4Ω
Frequency 0 to 50000.0 Hz 0.1 Hz 0.005%RD + 1 Hz
Pulse 0 to 999999 1 N/A
DC24V
(MAX 50mA)
24 V N/A 0.5 V
Environmental Specifications
Operating Temperature -10°C to 50°C
Storage Temperature -20°C to 60°C
Humidity <90%, non-condensing
Safety Specifications
European Compliance CE Mark
Mechanical Specifications
Display 3.5 inch TFT color screen
Electrical Connection Ø4mm sockets and at mini-jack thermocouple socket
RS232 Interface Standard RS232-DB9 socket
Size 3.9" x 7.6" x 2.0" (100 mm x 192 mm x 52 mm)
Weight 1.6 lb (0.7 kg)
Power Supply Polymer Li-ion rechargeable battery,or 10 V DC adapter
Battery Rechargeable Li-ion battery (included)
Battery Life
15 hours uninterrupted use
Battery life will be reduced when 24 V is applied
Battery Charge 110 V/220 V external power adapter (included)
Pressure Specification(222A & 223A)
The 160A series Intelligent Digital Pressure Modules are available for
gauge, vacuum and absolute pressure from -15 psi to 10,000 psi (-1 bar to
700 bar). Accuracy from 0.02% FS includes operation over 14°F to 122°F
(-10°C to 50°C), one year stability and calibration uncertainty. For detail
specication refer to pressure modules datasheet.
Built-in temperature readout
CVD coefcients of a calibrated PRT can be input into
the calibrator for accurate temperature measurement.
Multi lingual interface
English, German, French, Italian, Spanish,
Portuguese, Simplied Chinese
(Traditional Chinese, Japanese and Russian are
available per request)
Build-in unit conversion tool
Build-in converters for pressure units, temperature
units, temperature vs. resistance (RTDs), and
temperature vs millivolt (thermocouples)
Warranty: 3 years
HART Communication capability (223A)
Support HART
®
instrumentation
Display
3.5 inch TFT color screen
Misuse protection
Up to 30V voltage on any two sockets and up to
1A current on current sockets will not damage the
calibrator. The calibrator will return to normal condition
as soon as the voltage or current is removed.
Rechargeable battery
Rechargeable Li-ion battery for 15 hours uninterrupted
use.
Battery life will be reduced when 24 V is applied.
The rechargeable battery is replaceable.
*Accuracy with external cold junction; for internal cold junction add 0.1°C (k=2)
Pressure / Process Calibration Equipment
03
Corporate Headquarters
2900 Saturn St #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Fax: 714-998-6999 Email: sales@additel.com
Rev # 20210616
SPECIFICATIONS
ORDERING INFORMATION
Measurement Accuracy
Measure and Simulate
Standard Temperature Range (°C)
Accuracy (°C)
Measure (2W/3W) Measure (4W) Source
Pt10(385) IEC 751 -200 to 850
-200 to 200 0.65 0.60 0.65
200 to 600 0.82 0.72 0.82
600 to 850 0.96 0.82
0.96
Pt100(385)
IEC 751 -200 to 850
-200 to 200 0.15 0.1
0.15
200 to 600 0.26 0.16 0.26
600 to 850 0.34 0.20 0.34
Pt100(3916) JIS 1604 -200 to 850
-200 to 200 0.15 0.1 0.15
200 to 600 0.26 0.16 0.26
600 to 850 0.33 0.20 0.33
Pt100(3926) Minco Application Aid #18 -200 to 850
-200 to 200 0.15 0.1 0.15
200 to 600 0.26 0.16 0.26
600 to 850 0.33 0.20 0.33
Pt500(385) IEC 751 -200 to 850
-200 to 200 0.20 0.16 0.36
200 to 600 0.32 0.22 0.54
600 to 850
0.40 0.27
0.67
Pt1000(385) IEC 751
-200 to 850
-200 to 200 0.1
0.05 0.25
200 to 600 0.2 0.10 0.42
600 to 850 0.27 0.14 0.54
Cu10(427) Minco Application Aid #18 -100 to 260 -100 to 260 0.61 0.56 0.61
Cu50(428) GOST 6651-94 -50 to 150 -50 to 150 0.17 0.13 0.17
Cu100(428) GOST 6651-94 -50 to 150 -50 to 150 0.12 0.09
0.12
Ni120(672) Edison curve #7 -100 to 260 -100 to 260 0.07 0.05
0.07
Ni100(618) DIN 43760 -100 to 260 -100 to 260 0.08 0.06 0.08
* Additel/Land software can be downloaded for free at www.additel.com
Model Number
Accessories (included)
9816-X 110V/220V external power adapter 1 pc
9712 Chargeable Li-ion battery 1 pc
9022 Test leads 3 sets(6 pcs)
9020 Short circuit cable 1 sets(2 pcs)
9060
Pressure module connection cable
(only for 222A & 223A)
1 pc
Manual 1 pc
ISO 17025 accredited calibration
certicate
1 pc
Optional Accessories
Model number Description
ADT160A
(only for 222A & 223A)
Intelligent Digital Pressure Modules, see page 27
9060
(only for 222A & 223A)
Pressure module connection cable
9050 USB to RS232 (DB-9 Male) Adapter
9050-EXT RS 232 (DB9/M) extension cable, 9 feet
9080
Cable kit (including TC plug, compensation cable,
S,R,B,K,J,T,E,N)
9712
Spare chargeable Li-ion battery for multifunction
calibrator
9816
110 V/220 V external power adapter for ADT22X
and ADT672 calibrator
9906 Carrying case for multifunction calibrator
9510
Additel/Cal Task management software for
multifunction calibrator
ADT222A
ADT223A
ADT221A