Pressure / Process Calibration Equipment
01
Corporate Headquarters
2900 Saturn St #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Email: sales@additel.com
Rev # 20240109
Multifunction Process Calibrator
Additel 226, 226Ex
Sourcing, Simulating and Measuring Pressure,
Temperature and Electrical Signals
Built-in Barometer
Intrinsically Safe Models Available (Ex)
Large Smartphone Like Touchscreen User Experience
USB Type-C and Bluetooth Communications
IP67 Rated
High Voltage Measurement Capability (300V AC)
True RMS Voltage Meter Capability
Dual Channel Pressure Module Ports
High Static Differential Pressure Measurement 0.002% FS
ISO 17025-accredited Calibration w/data Included
OVERVIEW
Additel’s new Multi-functional Process Calibrator series takes portability, functionality, and accuracy to a whole new level
and packages it with an intuitive and easy to use color touchscreen display. The ADT226 is a powerful yet cost effective
process calibrator, which has an ATEX certied intrinsically safe option - ADT226Ex allowing you to perform calibration
work in the harshest of environments. We’re condent these new tools will not only meet your calibration requirements
but will make metrology simple for you!
Features
Easy-to-use Cellphone Like Interface
Additel's new and improved ADT226 series provides much improved
accuracies including an electrical accuracy of 0.015% RD + 0.005% FS,
high-static differential pressure mode accuracy to 0.002% FS and across
the board improvements in temperature measurement accuracies.
The ADT226 series brings an all new user interface to the world of process
calibrators. With a menu driven interface and small size/weight, the ADT226
is the industry's smallest multifunctional process calibrator with an intrinsically
safe version to boot (ADT226Ex).
It adopts advanced human hand engineering design for the most convenient
eld handheld process calibrator available. The ADT226 has been developed
with a powerful embedded operating system which solves common problems
of other designs including slow response, cumbersome key operation, high
power consumption and overall slow processing.
Accuracy
The ADT226 series delivers highly improved thermocouple measurement
capabilities by vastly improving the cold junction compensation(CJC)
specications and a much improved stabilization time.
Thermocouple Measurement Performance
Pressure / Process Calibration Equipment
02
Corporate Headquarters
2900 Saturn St #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Email: sales@additel.com
Rev # 20240109
The demands of remote calibration work can be challenging. The ADT226 series is
lightweight and highly portable and utilizes an advanced color LCD screen to help ensure
you can easily see, even in the (Ex) intrinsically safe versions.
All models in the ADT226 family have been designed with ruggedness and dependability in
mind and meet IP67 standards with a 1-meter drop test, 4G vibration, xenon exposure and
130g steel ball drop testing of the display.
Other environmental conditions have also been considered, such as temperature and
humidity. To combat these external elements, Additel has designed a unique internal circuit
design and process technology to allow for the utmost condence in your critical calibration
and measurement work.
The Additel 226Ex series calibrators have passed the most stringent testing by
certified organizations to acquire intrinsically safe certificates, ATEX, IECEX.
The explosion-proof grade (Ex ia IIC T4 Ga), can be widely used in potentially
explosive environments, such as oil and gas platforms, oil refineries, chemical
and petrochemical plants, pharmaceutical industries, energy and gas processing
industries.
Each intrinsically safe calibrator has an advanced transective color LCD display
which has enhanced visibility when viewed in direct sunlight. No matter where your
work takes you, these calibrators are up to the task.
The Additel 226 non-Ex version is equipped with “true effective
value” RMS measuring function, which can measure the RMS
of various waveforms with no need to consider distortion or
waveform parameters and other errors caused by various
waveforms
The onboard applications provide a useful selection of features including high static
differential pressure mode, pressure leak test, safety valve test, analog transmitter
calibration, unit converter, thermal calculator, and snapshots to name a few.
High static differential pressure mode uses two sensors, unique calculation technology to
achieve a differential pressure measurement to 0.002% FS at high static pressures. The
leak test will automatically calculate the pressure drop to determine a leak condition. The
safety valve test is a specialized task which captures the exact pressure release point by
taking 10 readings per second during a valve crack test.
You will nd this and much more as we continue to develop new apps at Additel.
Features
Portable and Robust
Intrinsically Safe Option
Voltage Meter (RMS)
Targeted Application Features
Users can remotely connect mobile devices to the ADT226 via Bluetooth with an
unobstructed distance up 20 meters. The included USB type-C comm port and cable
provide a hard wired communication option as well as charging for the removeable Li-ion
battery, which provides up to 12 hours of run time.
In addition to all the great features mentioned above, the ADT226 series is loaded with time saving
features like our builtin pressure and temperature converter, thermal calculator, wiring diagram
guide for assisting with electrical connections, a built-in diagnostic center including intelligent alarm
messaging and a real time error report and comprehensive selftesting to help our customers get the
very most out of their investment in Additel calibration tools.
Connectivity & Battery
Time Saving Features
Pressure / Process Calibration Equipment
03
Corporate Headquarters
2900 Saturn St #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Email: sales@additel.com
Rev # 20240109
SPECIFICATIONS
Source Accuracy
Specifications
ADT226 ADT226Ex
Range Resolution Accuracy Range Resolution Accuracy
Voltage DC
-150 to 150 mV 5 uV 0.015%RDG + 15 uV
0 to 10.5 V 0.2 mV 0.02%RDG + 0.5 mV
-1.5 to 1.5 V 0.05 mV 0.015%RDG + 0.15 mV
-15 to 15 V 0.5 mV 0.015%RDG + 1.5 mV
Current DC 0 to 25 mA 0.5 uA 0.015%RDG + 1.2 uA 0 to 25 mA 0.5 uA 0.02%RDG + 1.2 uA
Resistance
0 to 400 Ω 10 mΩ 0.015%RDG + 20 mΩ 0 to 400 Ω 10 mΩ 0.02%RDG + 20 mΩ
0 to 4000 Ω 100 mΩ 0.015%RDG + 200 mΩ 0 to 4000 Ω 100mΩ 0.02%RDG + 200 mΩ
Frequency
(Square wave)
(0.01 ~ 5) Hz 0.00001 Hz 0.005%RDG + 0.00005 Hz (0.01 ~ 5) Hz 0.00001 Hz 0.005%RDG + 0.00005 Hz
(5 ~ 50) Hz 0.0001 Hz 0.005%RDG + 0.0005 Hz (5 ~ 50) Hz 0.0001 Hz 0.005%RDG + 0.0005 Hz
(50 ~ 500) Hz 0.001 Hz 0.005%RDG + 0.005 Hz (50 ~ 500) Hz 0.001 Hz 0.005%RDG + 0.005 Hz
(500 ~ 5000) Hz 0.01 Hz
0.005%RDG + 0.05 Hz
(500 ~ 5000) Hz 0.01 Hz
0.005%RDG + 0.05 Hz
5000 ~ 50000) Hz 0.1 Hz 0.005%RDG + 0.5 Hz
5000 ~ 50000) Hz 0.1 Hz 0.005%RDG + 0.5 Hz
Frequency
(Sine wave &
Triangular wave)
(0.1~ 50) Hz 0.001 Hz 0.004 Hz
N/A
(50 ~ 500) Hz 0.01 Hz 0.04 Hz
(500 ~ 5000) Hz 0.1 Hz 0.4 Hz
(5000 ~ 50000) Hz 1Hz 4 Hz
Duty Cycle
(1%~99%)@≤10000Hz 0.05%
0.1% / kHz + 0.1% Fixed 50%@(0.01~50000)Hz
(5%~99%)@≤50000Hz 0.5%
Voltage mV (TC) -10 to 75 mV 1.5 uV 0.015%RDG + 4.0 uV -10 to 75 mV 1.5 uV 0.02%RDG + 4.0 uV
Pulse
0 to 9999999 1 N/A 0 to 9999999 1 N/A
Optional rising edge and falling edge, minimum threshold voltage: 2.5V
Loop power (max
25mA)
24 V N/A ±1 V 22 V N/A ± 10%
Electrical Specification
Note 1: When the environment temperature is (-10
+10)
and (30
50)
, the temperature coefcient is:
Voltage, current, thermocouple, thermal resistance output: ± 5 ppm FS/
(for Non-Ex version);
When the environment temperature is (-20
-10)
, the temperature coefcient is:
Voltage, current, thermocouple, thermal resistance output: ± 5 ppm FS/
(for Ex version);
Note 2: Output features:
Voltage output : ±150 mV /±1.5V / ± 15V, Maximum load current: 10 mA, (For Ex-version load current 5mA), load effect: 50 uV / mA;
Current output (0
25) mA: Maximum open circuit voltage: 24 V, driving capacity: 1 kΩ / 20 mA, maximum external voltage: 50 V;
(For Ex-version,Maximum open circuit voltage: 15 V, impedance: 400Ω, driving capacity: 6 V / 20 mA, maximum external voltage: 30 V
Frequency output: square wave, adjustable duty cycle, square wave amplitude (0~15) V adjustable, amplitude accuracy ± 0.2%FS(for Non-Ex version);
Frequency output: square wave, 50% duty cycle, square wave amplitude (0~10.5) V adjustable, amplitude accuracy ± 0.2%FS(for Ex version);
maximum load current: 10mA ( For Ex-version,1mA);
Supported units: Hz, kHz, MHz, CPM, CPH, s, ms ;
Zero-crossing sine wave / triangular wave amplitude: (0.1
30) Vp-p adjustable(only for Non-Ex version),
Amplitude accuracy 3 % Vp-p + 75 mV, supporting display valid value.
[1]
Thermocouple output: maximum load current: 5mA, load effect: < 5 uV / mA;
Thermal resistance output: maximum excitation current: Iex*400<1.6V(0 ~ 400) Ω, Iex*Rsim<1.6V(400 ~ 4000) Ω;
minimum excitation current: 0.2 mA@(0 ~ 400) Ω, 0.1 mA@(400 ~ 4000) Ω ;
support 1ms pulse excitation. (For Non-Ex version
Thermal resistance output: Excitation current: (0.2~2) mA@(0 ~ 400) Ω, (0.1~0.3) mA@(400 ~ 4000) Ω ;
support 1ms pulse excitation. (For Ex version
Pressure / Process Calibration Equipment
04
Corporate Headquarters
2900 Saturn St #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Email: sales@additel.com
Rev # 20240109
Measurement Accuracy Cont.
Specifications
ADT226 ADT226Ex
Range Resolution Accuracy Range Resolution Accuracy
Voltage DC
-300 to 300 mV 1 µV 0.015% RDG + 15 µV -300 to 300 mV 1µV
0.02% RDG + 15 µV
-30 to 30 V 0.1 mV 0.015%RDG + 1.5 mV -30 to 30 V 0.1 mV 0.02% RDG + 1.5 mV
Impedance: -300 mV to 300 mV = > 100 MΩ
-30 V to 30 V = >1 MΩ
DC High Voltage
-300 to 300 V 10 mV 0.05% RDG + 30 mV
N/A
The highest input voltage is 300 V, IEC61010 300V CATII
Common mode rejection: >100 dB (at 50 or 60 Hz)
Impedance: > 4 MΩ, DC coupling
AC High Voltage
300V (40 to 500 Hz) 10 mV 0.5% RDG + 150 mV
N/A
The highest input voltage is 300 V, IEC61010 300V CATII
9% to 100% of range is suitable for the above accuracy indicators
Impedance: >4 MΩ, <100pF, AC coupling
Current DC -30 to 30 mA 0.1 μA 0.015% RDG + 1.5 µA -30 to 30 mA 0.1 μA 0.02% RDG + 1.5 µA
Resistance
(4-Wire)
0 to 400 Ω 1 mΩ 0.015% RDG + 20 mΩ 0 to 400 Ω 1 mΩ 0.02% RDG + 20 mΩ
0 to 4000 Ω 10 mΩ 0.015% RDG + 200 mΩ 0 to 4000 Ω 10 mΩ 0.02% RDG + 200 mΩ
2-Wire + 50 mΩ, 3-wire + 10 mΩ
Excitation current: 0.2 mA
Voltage mV (TC)
-10 to 75 mV 0.1uV 0.015% RDG + 4.0 µV -10 to 75 mV 0.1uV
0.02% RDG + 4.0 µV
Impedance: >100 MΩ
Frequency
(0.01 ~ 5) Hz 0.00001 Hz 0.005%RDG + 0.00005 Hz (0.01 ~ 5) Hz 0.00001 Hz 0.005%RDG + 0.00005 Hz
(5 ~ 50) Hz 0.0001 Hz 0.005%RDG + 0.0005 Hz (5 ~ 50) Hz 0.0001 Hz 0.005%RDG + 0.0005 Hz
(50 ~ 500) Hz 0.001 Hz 0.005%RDG + 0.005 Hz (50 ~ 500) Hz 0.001 Hz 0.005%RDG + 0.005 Hz
(500 ~ 5000) Hz 0.01 Hz 0.005%RDG + 0.05 Hz (500 ~ 5000) Hz 0.01 Hz 0.005%RDG + 0.05 Hz
(5000 ~ 50000) Hz 0.1 Hz 0.005%RDG + 0.5 Hz (5000 ~ 50000) Hz 0.1 Hz 0.005%RDG + 0.5 Hz
Minimum threshold voltage: 2.5 V
Supported units: Hz, kHz, MHz, CPM, CPH, s, ms, µs
Duty Cycle
(1%~99%)@≤10000Hz 0.01%
0.1% kHz + 0.05% N/A
(5%~99%)@≤50000Hz 0.1%
Pulse
0 to 9999999 1 N/A 0 to 9999999 1
N/A
Optional rising edge and falling edge, minimum threshold voltage: 2.5V
Switch Support for dry or wet switch, voltage range of 3 to 30 V, response speed of < 10 ms
SPECIFICATIONS
Note 1: When the environment temperature is (-10
+10)
and (30
50)
, the temperature coefcient is:
Voltage, current, thermocouple, thermal resistance output: ± 5 ppm FS/
(for Non-Ex version);
When the environment temperature is (-20
-10)
, the temperature coefcient is:
Voltage, current, thermocouple, thermal resistance output: ± 5 ppm FS/
(for Ex version);
AC High Voltage TRMS measurement: ± (250 ppmRDG + 25 ppmFS)/
℃;
DC High Voltage measurement:± 25ppmFS/
℃.
Note 2: Input features:
Voltage range: (-300 ~ 300) mV, input impedance >100 MΩ; (-30 ~ 30) V, input impedance >1MΩ;
Current measurement: input impedance < 40 Ω;
TC measurement: input impedance >100 MΩ;
AC High Voltage TRMS measurement: input impedance: > 4MΩ , <100pF, AC coupling; Maximum input voltage: 300 V, IEC61010 300V CATII;
9% ~ 100% of the range is applicable to the accuracy index above.
DC High Voltage measurement: > 4 MΩ, DC coupling; Maximum input voltage: 300 V, IEC61010 300V CATII; Common-mode rejection:>100 dB
in 50 or 60 Hz
Note 3
:
The thermal resistance measurement excitation power supply is 0.2mA. There are four wire system, three wire system and two wire system measurement modes
at each gear position. The accuracy indicators are as follows
The accuracy data given in the table is the accuracy data in 4-line system
3-wire system accuracy is +10 mΩ on the basis of 4-wire system accuracy;
2-wire accuracy is +50 mΩ on the basis of 4-wire accuracy;
Note 4: Minimum threshold voltage for frequency and pulse measurement: 2.5V;
Note 5: Frequency measurement unit: Hz, kHz, MHz, CPM, CPH, s, ms, µs;
Note 6: Optional rising edge or descending edge trigger mode for pulse measurement.
Pressure / Process Calibration Equipment
05
Corporate Headquarters
2900 Saturn St #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Email: sales@additel.com
Rev # 20240109
Specifications
ADT226 ADT226Ex
Operating Temperature -10°C to 50°C -20°C to 50°C
Specification guaranteed
temperature range
10°C to 30°C -10°C to 50°C
Storage Temperature
-30°C to 70°C -30°C to 70°C
Humidity <95%, non-condensing <95%, non-condensing
Power supply
6600mAh, 23.8Wh lithium battery, charging time about 6
hours, battery pack can be charged independently
4000mAh 14.4Wh Explosion-proof lithium battery packcharging time
about 6 hours, battery pack can be charged independently
User interface
Icon drive menus Icon driven menus with navigation buttons
Ports protection voltage
50V max
Only for the top ports)
30V max
Display 5.0 inch 480 x 800 mm TFT LCD capacitive screen 4.4 inch 640 x 480 mm color display capacitive screen
Maximum altitude 3000 meters
European Compliance CE Mark
Electrical Connection Ø4mm sockets and at mini-jack thermocouple socket
Size 6.97" x 4.13" x 2.04" (177 mm x 105 mm x 52 mm)
Weight 1.6 lb (0.7 kg) 1.65 lb (0.75Kg)
Battery Rechargeable Li-ion battery (included)
Battery Life Typically 12 hours Typically 35 hours
Battery Charge
110V/220V external power adapter included. Battery can be charged external to the unit.
External pressure module Dual channel Serial plug, can connect two digital pressure modules
Warm-up time Full specication performance is achieved after a 10 minute warm-up time.
ROHS compliant Rohs II Directive 2011/65/EU, EN50581:2012
Display rate 3 readings per second
Barometric Accuracy
(Built-in barometer)
55Pa
IP protection level IP67, 1 meter drop test
Communication Isolate USB-TYPEC (slave), Bluetooth BLE
User Interface
Localization
English, German, French, Italian, Spanish, Portuguese,
Simplied Chinese, Traditional Chinese, Japanese,
Russian, Czech, Slovak
English, Simplied Chinese, Traditional Chinese, Japanese
Calibration
ISO 17025 accredited calibration with data
Warranty 3 years
General Specification
Pressure / Process Calibration Equipment
06
Corporate Headquarters
2900 Saturn St #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Email: sales@additel.com
Rev # 20240109
Temperature Specification
SPECIFICATIONS
Thermocouple Measurement and Source Accuracy
ADT226 ADT226Ex
Type Standard Temperature Range (
o
C)
Accuracy (
o
C)
Standard Temperature Range (
o
C)
Accuracy (
o
C)
Measure / Source Measure / Source
S IEC 584 -50 to 1768
-50~0 0.96
IEC 584 -50 to 1768
-50~100 0.96
0~100 0.69 100~1000 0.69
100~1768 0.64 1000~1768 0.73
R IEC 584 -50 to 1768
-50~0
1.02
IEC 584 -50 to 1768
-50~0 1.03
0~200 0.71 0~200 0.71
200~1768 0.56 200~1768 0.65
B IEC 584 0 to 1820
200~300 1.89
IEC 584 0 to 1820
200~300 1.90
300~500
1.25 300~500 1.26
500~800 0.78 500~800 0.79
800~1820 0.55 800~1820 0.57
K
IEC 584
-270 to 1372
-250 to -200 0.97
IEC 584
-270 to 1372
-250 to -200 1.04
-200 to -100 0.30 -200 to -100 0.32
-100 to 600 0.18 -100 to 600 0.21
600 to 1372 0.35 600 to 1372 0.43
N IEC 584 -270 to 1300
-250 to -200 1.50
IEC 584 -270 to 1300
-250 to -200 1.58
-200 to -100 0.44 -200 to -100 0.46
-100 to 1300 0.30 -100 to 1300 0.37
E IEC 584 -270 to 1000
-250~-200 0.54
IEC 584 -270 to 1000
-250~-200 0.59
-200~-100 0.20 -200~-100 0.22
-100~700 0.15 -100~700 0.18
700~1000 0.20 700~1000 0.25
J
IEC 584 -210~1200
-210~-100 0.26
IEC 584 -210~1200
-210~-100 0.28
-100~700 0.15 -100~700 0.19
700~1200 0.25 700~1200 0.31
T IEC 584 -270 to 400
-250~-100 0.74
IEC 584 -270 to 400
-250~-100 0.79
-100~0 0.15 -100~0 0.16
0~400 0.11 0~400 0.13
C ASTM E988 0 to 2315
0 to 1000 0.35
ASTM E988 0 to 2315
0 to 1000 0.40
1000 to 1800 0.62 1000 to 1800 0.73
1800 to 2315 1.02 1800 to 2315 1.22
D ASTM E988 0~2315
0~100 0.39
ASTM E988 0~2315
0~100 0.39
100~1200 0.37 100~1200 0.43
1200~2000 0.65 1200~2000 0.77
2000~2315 1.03 2000~2315 1.24
G ASTM E1751 0 to 2315
50~100 1.12
ASTM E1751 0 to 2315
50~100 1.12
100~200 0.72 100~200 0.72
200~400 0.45 200~400 0.46
400~1500 0.37 400~1500 0.43
1500~2315 0.77 1500~2315 0.92
L DIN43710 -200 to 900
-200 to -100 0.15
DIN43710 -200 to 900
-200 to -100 0.16
-100 to 400 0.13 -100 to 400 0.14
400 to 900 0.17 400 to 900 0.20
U DIN43710 -200 to 600
-200 to 0 0.28
DIN43710 -200 to 600
-200 to 0 0.29
0 to 600 0.13 0 to 600 0.15
Note: Internal CJC is ±0.2°C (-10°C to 50°C ambient temperature)
Accuracy with external cold junction only, for internal cold junction add 0.2°C (k=2)
Pressure Specification( ADT226 & ADT226Ex)
The 161 series Intelligent Digital Pressure Modules are available for gauge, vacuum and absolute pressure from -15 psi to 60,000 psi (-1 bar to 4200 bar).
Accuracy from 0.02% FS includes operation over 14°F to 122°F (-10°C to 50°C), one year stability and calibration uncertainty. For detailed specications,
please refer to the pressure modules datasheet.
Pressure Specification
Pressure / Process Calibration Equipment
07
Corporate Headquarters
2900 Saturn St #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Email: sales@additel.com
Rev # 20240109
RTD Measurement and Source Accuracy
Measure and Simulate Temperature Range (°C)
Accuracy (°C)
ADT226 ADT226Ex
PT10(385) -200 to 850
-200~200 0.62 0.64
200~600 0.77 0.82
600~850 0.88 0.95
PT25(385) -200 to 850
-200~200 0.29 0.31
200~600 0.40 0.44
600~850 0.47 0.54
PT50(3916) -200 to 850
-200~200 0.18 0.20
200~600 0.27 0.32
600~850
0.34
0.40
PT100(385)
PT100(391)
PT100(3916)
PT100(3926)
-200 to 850
-200~200 0.13 0.15
200~600 0.21
0.26
600~850 0.27 0.34
PT200(385) -200 to 850
-200~200 0.34 0.37
200~300
0.37 0.40
300~600 0.46 0.51
600~850 0.54 0.61
PT400(385) -200 to 850
-200~0 0.17 0.18
0~200 0.21 0.23
200~600 0.30 0.35
600~850 0.37 0.44
PT500(385) -200 to 850
-200~200 0.18 0.20
200~600 0.27 0.32
600~850 0.34 0.40
PT1000(385) -200 to 850
-200~200 0.13 0.15
200~600 0.21 0.26
600~850 0.27 0.34
Cu10(427) -200~260 -200~260 0.59 0.61
Cu50(428) --200~260 -200~260 0.15 0.17
Cu100(428) -200~260 -200~260 0.10 0.12
Ni100(617)
Ni100(618)
-60~180
-60~0 0.06 0.07
0~180 0.06 0.08
Ni120(672) --80~260 -80~260 0.06 0.07
Ni1000 -50~150 -50~150 0.08 0.09
*Note: Ambient temperature of 20°C±10°C.
4-wire accuracy. For 2-wire add 50 mΩ, for 3-wire add 10 mΩ
SPECIFICATIONS
Pressure / Process Calibration Equipment
08
Corporate Headquarters
2900 Saturn St #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Email: sales@additel.com
Rev # 20240109
ORDERING INFORMATION
* Additel/Land software can be downloaded for free at www.additel.com
Model Number
Accessories (included)
Model number Description QTY Picture
9811-X
110V/220V external power adapter
(Only for ADT226
,
226P)
1 pc
9811Ex-X
110V/220V external power adapter
(Only for ADT226Ex )
1 pc
9704
Chargeable Li-ion battery
(Only for ADT226, 226P)
1 pc
9704Ex
Chargeable Li-ion battery
(Only for ADT226Ex )
1 pc
9023 Test leads
1 set
(6 pcs)
9027
Right angle test leads
(Non-Ex models only)
1 set
(2 pcs)
9060 Pressure module connection cable 1 pc
9052
USB Cable type A to type C
(Non-Ex models only)
1 pc
9052Ex
Ex USB Cable type A to type C
(For Ex models only)
1 pc
9040 Hanging strap with magnet 1 pc
9028
Multimeter Test Hook, Flexible Electronic
Probe
1 set
(2 pcs)
ISO 17025 accredited calibration certicate 1 pc
ADT226
ADT226
ADT226Ex: Intrisically Safe
ADT226P: Panel Mount
Optional Accessories
Model number Description Picture
Model number Description Picture
ADT161 - XXX Digital Pressure Modules 9082
HART 250 ohm resistor adapter for
ADT226, 226Pand ADT227, 227P
ADT161Ex - XXX
Intrinsically Safe Digital Pressure
Modules
9704
Battery, rechargeable Li-ion polymer
battery for Additel Handheld Series
ADT129-X
Differential Pressure Manifold, -15 to
3,000 psi
9704Ex
Battery, rechargeable Li-ion polymer
battery for Ex Additel Handheld Series
9061
Current output cable (for ADT226,
226P and ADT227, 227P)
9811-X
110 V/220 V external power adapter
for handheld models
9062
Connection adapter cable for Fluke
style pressure modules to non-
explosion-proof Additel readouts
9811Ex-X
110 V/220 V external power adapter
for Ex handheld models
AM1602-6FT
Class A, PT100/385 Industrial RTD,
-40°C to 160°C, 3/16 (4.76 mm)
inch x 2 inch (50 mm) with 6 foot
(1.8 Meters) cable w/ banana jack
connectors
9906A
Hard carrying case for handheld
instrument with accessories
9080
Cable kit (including TC plug,
compensation cable, S,R,K,J,T,E,N)
9918-SC
Soft carrying case, with space for
handheld instrument, test leads, and
accessories
9081
Universal TC easy-press adapter for
ADT226, 226P and ADT227, 227P
9530-BASIC
Additel/Acal Task management
software for multifunction calibrator
9530-NET
Additel/Acal Automated calibration
software with asset management,
network version, Includes server
installation and 1 user license
Panel Mount Version