Specication Range Accuracy Resolution Note
RTD Measurement
Accuracy
0~400 Ω 0.01%RDG + 20 mΩ
[1]
1 mΩ Excitation current: 1 mA
Voltage Measurement
±300 mV 0.015%RDG + 1.5 mV 1 uV Impedance: >100 MΩ
±30 V 0.015%RDG + 1.5 μA 0.1 mV Impedance: >1 MΩ
Current Measurement ±30 mA 0.015%RDG + 1.5 μA 0.1 uA Impedance: < 40 Ω
Frequency Measurement
0.01~50000Hz (auto range) 0.005% RDG + 2 last digit 6-digit auto-resolution Min threshold voltage: 2.5V
Units: Hz, kHz, MHz, CPM, CPH, s, ms, us
Switch On-Off
Measurement
Inspection voltage: (3 ~ 30)V
Response speed: < 10ms
,
supports wet and dry switch
Pulse Count
0 ~ 9999999, optional rising edge and falling edge
Min threshold voltage: 2.5 V
Loop Power 22 V ± 10%, max output impedance: 320 Ω, max load current: 25 mA
Electrical Specifications
PRESSURE TECHNICAL SPECIFICATIONS
Specications
Pressure resolution 4, 5, or 6 digit resolution (user selectable)
Measurement Units
[1]
kPa, Pa, GPa, MPa, mPa, uPa, hPa, bar, mbar, torr, mtorr, atm, psi, psia, psig, gf/cm2, kgf/cm2, inH2O@4°C, inH2O@68°F,
mmH2O@4°C, mmH2O@20°C, ftH2O@4°C, ftH2O@68°F, inHg@0°C, mmHg@0°C, lb/ft2, tsi, psf, inH2O@60°F,
ftH2O@60°F, cmH2O@4°C, mH2O@4°C, cmHg@0°C, mHg@0°C, kgf/m2
Temperature Compensation
-10°C
~
50 °C
Pressure Module Type
Built-in digital pressure module: ADT158Ex, for more detailed information, please see ADT 158Ex datasheet.
External digital pressure module: ADT161Ex,for more detailed information, please see ADT 161Ex datasheet.
Specications Refer to the technical specication of the ADT158Ex and ADT161Ex.
High Static Pressure and
Differential Pressure
Synthesis Index
[2]
Two modules must be with the same range;
Typical Differential pressure accuracy of 0.002%FS or 0.02%RD, whichever is greater when using two 0.02%FS external
modules.
Typical Differential pressure accuracy of 0.002%FS or 0.05%RD, whichever is greater when using two 0.05%FS external
modules.
[1] Available units are dependent on the overall pressure range.
[2] For more explanation, please reference application note "Achieving High Accuracy for High Static Differential Pressure Measurements".
Pressure Ranges
Gauge Pressure
[1]
P/N
Pressure Range
Media
[2]
Accuracy (%FS)
Burst
Pressure
(psi) (bar)
V15 -15 -1.0 G 0.02 3×
GP2 2 0.16 G 0.05 3×
GP5 5 0.35 G 0.05 3×
GP10 10 0.7 G 0.02 3×
GP15 15 1.0 G 0.02 3×
GP30 30 2.0 G 0.02 3×
GP50 50 3.5 G , L 0.02 3×
GP100 100 7.0 G , L 0.02 3×
GP150 150 10 G , L 0.02 3×
GP300 300 20 G , L 0.02 3×
GP500 500 35 G , L 0.02 3×
GP600 600 40 G , L 0.02 3×
GP1K 1,000 70 G , L 0.02 3×
GP1.5K 1,500 100 G , L 0.02 3×
GP2K 2,000 140 G , L 0.02 3×
GP3K 3,000 200 G , L 0.02 3×
GP5K 5,000 350 G , L 0.02 3×
GP10K 10,000 700 G , L 0.02 2×
GP15K 15,000 1,000 G , L 0.05 2x
GP20K 20,000 1,400 G , L 0.05 1.5x
GP25K 25,000 1,600 G , L 0.05 1.5x
GP30K 30,000 2,000 G , L 0.05 1.5x
GP36K 36,000 2,500 G , L 0.05 1.5x
GP40K 40,000 2,800 G , L 0.05 1.35x
GP50K 50,000 3,500 G , L 0.1 1.2x
GP60K 60,000 4,200 G , L 0.1 1.1x
Compound Pressure
P/N
Pressure Range
Media Accuracy
Burst
Pressure
psi bar
CP2
±2 ±0.16 G 0.05% FS 3x
CP5
±5 ±0.35 G 0.02% FS 3x
CP10
±10 ±0.7 G 0.02% FS 3x
CP15
±15 ±1.0 G 0.02% FS 3x
CP30
-15 to 30 -1 to 2.0 G 0.02% FS 3x
CP50
-15 to 50 -1 to 3.5 G 0.02% FS 3x
CP100
-15 to 100 -1 to 7.0 G,L 0.02% FS 3x
CP300
-15 to 300 -1 to 20 G,L 0.02% FS 3x
CP500
-15 to 500 -1 to 35 G,L 0.02% FS 3x
CP600
-15 to 600 -1 to 40 G,L 0.02% FS 3x
CP1K
-15 to 1,000 -1 to 70 G,L 0.02% FS 3x
CP2K
-15 to 2,000 -1 to 140 G,L 0.02% FS 3x
CP3K
-15 to 3,000 -1 to 200 G,L 0.02% FS 3x
CP5K
-15 to 5,000 -1 to 350 G,L 0.02% FS 3x
CP10K
-15 to 10,000 -1 to 700 G,L 0.02% FS 2x
[1] Sealed gauge pressure for above 1000 psi
[2] G=Gas, L=Liquid
Note [1]: Accuracy applies to 4-wire probes. For 3-wire probes add 10 mΩ, for 2-wire probes add 50 mΩ