Temperature -Humandity Recording Equipment
02
Pressure Calibration Equipment
General Specifications
Specification Description
Display 10.1 inch TFT color capacitive touch screen
Display Rate 10 readings per second
Resolution 5, 6, 7 digits (user selectable)
Size 11 × 9 × 6.26 in (281 × 230 × 159 mm)
Weight 10.48 lbs (4.8 Kgs) with 3 built-in modules and without external module and battery
Power Supply 28V DC adaptor, optional rechargeable Li-ion battery, typically 8 hours of operation
Units
psi, psf, atm, inH2O (4 °C), inH2O (20 °C), inH2O (60 °F), mbar, bar, Dy/cm2, pascal, hPa,kPa, MPa, inHg (0 °C), inHg
(60 °F), mTorr, Torr, mmHg (0 °C), cmHg (0 °C), mHg (0 °C), mmH2O (4 °C), cmH2O (4 °C), mH2O (4 °C), mmH2O (20
°C), cmH2O (20 °C), mH2O (20 °C), ftH2O (4 °C), ftH2O (20 °C), ftH2O (60 °F), tsi, tsf, g/cm², kg/cm², kg/m²
5 customizable units
2 height units: Meters and feet
Environment
Working temperature: 0 ~ 40 °C
Ensuring accuracy temperature:15 ~ 35 °C
Operating humidity: 0 % ~ 90% RH (non-condensing)
Storage temperature: -20 ~ 70 °C
Altitude:
<
2000 m
Compliance
CE
Warm-Up Time 20 minutes (for optimal performance)
Communication Wi-Fi (optional), USB*2, Ethernet*2, HDMI*1
Installation Methods
Supports desktop and rack installation
(screw posts are reserved at the bottom of the device, which can be installed into the cabinet)
Built-In Pressure Modules
(CH1 & CH2 & CH3)
(See built-in pressure modules
selection table)
Pressure port: 1/8 BSF M
Pressure ≤ 10K PSI: pre-installed with 1/8 BSP M to 1/4 BSP F adapter,
optional high-pressure hose connection with various general adapters.
Pressure > 10K PSI: pre-installed with 1/8 BSP M to 1/4HP M adapter,
optional ultra-high-pressure hose connection with various special adapters.
CH1 optional barometric pressure module;
CH2 and CH3 can be equipped with 1 or 2 built-in pressure modules selected by the user
External Module (CH4 & CH5)
(See ADT161 Selection Table)
5-pin Lemo connector for connection with external pressure modules
Accuracy
Barometric module: ±10 Pa (For BPH), ±22 Pa (For BP); Gauge module: 0.01%FS, 0.02%FS, 0.01% rdg or 0.005% FS
whichever is greater
Annual Stability 0.005%FS
Pressure Medium
≤ 15 psi: Dry, clean and non-corrosive gas compatible with 316SS
>
15 psi: Non-corrosive gas or liquid compatible with 316SS
Warranty 1 Year