Pressure / Process Calibration Equipment
03
Corporate Headquarters
2900 Saturn Street #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Email: sales@additel.com
Rev # 20260317
FEATURES
Pressure Specifications
Electrical Specifications (ADT760A)
ADT760A/760S-LLP
The 760A/760S-LLP is designed for low pressure calibration and comes with a build-in pressure module of your choice.
The maximum range module compatible with the 760A/760S-LLP is to ±30 inH2O (±75 mbar) and provides an accuracy
to 0.05%FS (see ordering information for congurations with the option of the 760A/760S and a module of your choice).
Additional internal pressure modules (ADT155 series) are available and provide a variety of ranges down to ±0.25
inH
2
O (±0.62 mbar). The accuracy of 0.05%FS and control stability 0.005%FS is based on the internal module’s span.
Measurements can be made in gauge or differential mode.
ADT760A/760S-D
The760A/760S-D gives you differential and gauge pressure but at a higher pressure range than the 760A/760S-LLP.
Covering the range of -12.5 to 35 psi (-0.86 to 2.5 bar) and with an accuracy of 0.01%FS or 0.02%FS depending on
the interchangeable ADT155 module used, the 760A/760S-D is an ideal solution to cover very
common gauge and differential pressure measurements. The Additel 760A/760S-D comes with
an internal module of your choice. The maximum range module compatible with the 760A/760S-D
is to 35 psi. Lower pressure congurations down to ±10 inH
2
O differential can be purchased to
improve accuracy at lower pressures.
ADT760A-MA
The 760A
/
760S-MA generates and controls pressure from vacuum pressures up to 375 psig (25
bar.g) with an accuracy of 0.01%FS or 0.02%FS depending on the interchangeable ADT155
module used
.
Equipped with a built-in barometric reference, each unit can switch between
gauge and absolute pressure types. A variety of internal sensors are available which offer lower
pressure ranges for improved performance.
Specification 760A/760S-LLP 760A/760S-D 760A/760S-MA
Max Pressure Range
±30 inH
2
O
(±75 mbar)
-12.5 to 35 psi
(-0.86 to 2.5 bar)
-12.5 to 375 psig
(-0.86 to 25 bar.g)
Accuracy 0.05%FS
[1]
0.01%FS or 0.02%
[1]
0.01%FS or 0.02%FS
[1][3]
Stability 0.005%FS
[2]
0.005%FS
[2]
0.005%FS
[2]
Pressure Type Differential, Gauge Differential, Gauge Gauge, Absolute
Over Range Indication 120%
Resolution 6 digits
Measurement Units
Pa, hPa, kPa, mPa, bar, mbar, psi, mmHg@0°C, cmHg@0°C, mHg@0°C, inHg@0°C, inH
2
O@4°C, mmH
2
O@4°C,
cmH
2
O@4°C, mH
2
O@4°C, mmH
2
O@20°C, cmH
2
O@20°C, mH
2
O@20°C, mH2O@68°F, inH
2
O@20°C, kgf/cm2, mtorr, torr, lb/
ft2, tsi, custom
Barometric Accuracy N/A N/A 55 Pa
[4]
Connection Barb tting
Hose, 5 ft (1.5 m), with built-in lter to
1/4BSPF, 1/4NPTF, and M20F adapters
Hose, 5 ft (1.5 m), with built-in lter to
1/4BSPF, 1/4NPTF, and M20F adapters
Pressure Output Rate <30 Seconds (30 inH
2
O / 100 ml) <10 Seconds (35 psi / 5 ml) <90 Seconds (300 psi / 5 ml)
Specification Range Resolution Accuracy Note
mA Measure ± 30 mA 0.0001 mA 0.01%RD + 1.5 μA Impedance <10 Ω
V Measure ± 30 V 0.0001 V 0.01%RD + 1.5 mV Impedance >1 MΩ
mA Source 24 mA 0.001 mA 0.01%RD + 1.2 μA 20 mA @ 1 K
Loop Power Source 24 V N/A ± 1 V 50 mA (Max Loading)
Pressure Switch Open, close. Support for mechanical switches and NPN/PNP digital switches.
Temperature Compensation 8 °C to 38 °C
Temperature Coefficient < ± ( 0.001%RD + 0.001%FS ) / °C outside of 8°C to 38°C
[1] FS specication applies to the span of the module range.
[2] Stability based on FS of the internal pressure module. Stability is 0.005%FS or 0.05 pa whichever is greater. Internal module is switchable.
[3] Specication based on gauge measurement. An additional 55 pa uncertainty will need to be included when measuring in absolute mode.
[4] 55 Pa uncertainty (k=2) includes calibration uncertainty, linearity, and long-term stability (<30 Pa per year). Barometer range of 60 to 110 kPa.