Automated Hydraulic Pressure Calibrator
Additel 762
OVERVIEW
The Additel 762 Automated Pressure Calibrator is unlike any other pressure calibrator
on the market. This revolutionary product is a complete turnkey solution for automation
of pressure calibration work up to 10,000 PSI.
Designed for use in both the eld and the laboratory, the portability and accuracy of this
state-of-the-art product will quickly become the favorite go-to calibrator for lab personnel
and eld technicians alike.
With fully automated support for calibration of pressure transmitters, switches, dial
and digital gauges and sensors, including HART/PROFIBUS devices in conjunction
with a fully integrated task feature, data collection and Wi-Fi connectivity, we had our
customer’s needs in mind when designing our most capable pressure calibrator to date.
Automated pressure generation and control to 10,000 psi (700 Bar)
Accuracy to 0.01%FS
Dual-range manual or auto select
Control stability 0.005%FS
Portable - designed for use in the field and in the lab
Control by optional external pressure modules
Supports two external pressure modules
Wi-Fi, LAN, Bluetooth, USB and Ethernet communication
Full HART field communicator
HART and PROFIBUS communication
Data logging and task management
Patented electric pump technology
Pressure Calibration Equipment
02
Corporate Headquarters
2900 Saturn St #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Email: sales@additel.com
Rev # 20230904
Dual-Range Accuracy to 0.01% FS
The ADT762 includes the unique ability to automatically switch between different
internal calibrations depending on the current control pressure of the ADT762.
Additel provides calibrations unique to each ADT762 for ranges of 0-3,000 PSI
(200 Bar) and 0-10,000 PSI (700 Bar). As the calibrator is pressurized, it will
automatically select the control and measurement specification based on the
specic pressure range. Pressure calibration range selection can be set to “auto”
mode so the calibration range is automatically selected by the ADT762 based on
the set point pressure, or the calibration range can be manually selected.
Dual-Mode HART Communication
HART pressure transmitters can be directly maintained and calibrated without any other equipment or tools.
The ADT762 provides an automatic HART calibration mode as well as a manual mode. This dual-mode HART
communication function not only provides an efcient and convenient interoperability mode for DUT, but also supports
access to a fully HART capable calibrator.
Documented Task Feature
The powerful documented calibration task application allows users to quickly
create and execute tasks without the need for a PC or tablet. The ADT762
automatically analyzes errors, generates test reports, while storing results locally.
The Additel 762 can support up to 1000 documented tasks which can be stored
and recalled at any time to help save time and money.
Built-in Auto-Purge Application
Purging hydraulic calibration systems can be challenging and time consuming.
The ADT762 has been designed with an integrated auto-purge system that saves
time, money and frustration by completely automating the removal of air from
the system. With the push of a button, the ADT762 quickly manages the system
purging. This helps to free up time for technicians to attend to other needs.
03
Corporate Headquarters
2900 Saturn St #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Pressure Calibration Equipment
Phone: 714-998-6899 Email: sales@additel.com
Rev # 20230904
Model
Specification
ADT762 Automated Hydraulic Pressure Calibrator
Pressure Range 15~10,000 psi (1-700 Bar)
Range Selection Manual 3K psi, Manual 10K psi or Auto-range
Accuracy
0~10,000 psi, 0.01%FS 0~10,000 psi, 0.02%FS
0~3,000 psi, 0.01%FS 0~3,000 psi, 0.02%FS
Resolution 7 digits 6 digits
Measurement Units
Pa, hPa, kPa, mPa, bar, mbar, psi, mmHg@0°C, cmHg@0°C, mHg@0°C, inHg@0°C, inH2O@4°C,
mmH2O@4°C, cmH2O@4°C, mH2O@4°C, mmH2O@20°C, cmH2O@20°C, mH2O@20°C,
inH2O@20°C, inH2O@68°F, kgf/cm2, mtorr, torr, lb/ft2, tsi, custom
Maximum External Load Capacity Max: 80 ml@700 Bar, 50 ml recommend
Reservoir Max: 350 ml
,
built-in lter
Control Stability
[1]
0.005%FS from 100 to 10,000 psi
Stability Duration
5 min
Pressure Module Built-in one module with dual range
External Control Pressure Module See the following "External Control Pressure Module Specication and Compatibility" table
External Measurement Pressure Module All ADT161 pressure modules
Pressure Specifications
Electrical Specifications
Supported
[1] Control Stability is based on the range selection or external module
Model
Specification
Range Resolution Accuracy Note
mA Measure
-25 to 25 mA 0.1 μA ±
(
0.008%RD + 1.0 μA
)
Impedance <10 Ω
-50 to 50 mA 0.1 μA ±
(
0.008%RD + 2.0 μA
)
V Measure -300 to 300 mV 1 μV ±
(
0.008%RD + 6 μV
)
Impedance <1 GΩ
V Measure
(Auto-ranging)
-5 to 5 V 20 μV ±
(
0.008%RD + 100 μV
)
Impedance >1 MΩ
-12 to 12 V 100 μV ±
(
0.008%RD + 320 μV
)
-30 to 30 V 100 μV ±
(
0.008%RD + 600 μV
)
Loop Power Source 24 V N/A ±1 V 50 mA (Max Loading)
mA Source
0 to 2.5 mA or
2.5 to 25 mA
0-2.5 mA: 0.05μA
0-25 mA:0.5μA
0-2.5 mA: 0.008%RD + 0.2 μA
0-25 mA: 0.008%RD + 1.0 μA
20 mA @ 1 KΩ
Power Source 16 to 30 V 1 V ±1 V 70 mA (Max Loading)
V Source 0 to 16 V 0.25 mV 0.008%RD + 500 μV
Pressure Switch
Mechanical Switch, Live
Mechanical Switch, NPN Switch,
PNP Switch
N/A N/A
Response time<10 ms.
If the switch is live, voltage range
will be (3-30) V
Temperature Compensation 18 °C to 28°C
Temperature Coefficient Outside of 18 °C to 28 °C: <± 0.0005%RD + 0.00005%FS/°C
Misuse Protection Up to 30 V on any two sockets
Pressure Switch Test
HART / PROFIBUS PA
Pressure Calibration Equipment
04
Corporate Headquarters
2900 Saturn St #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Email: sales@additel.com
Rev # 20230904
External Control Pressure Module Specification and Compatibility
General Specifications
Specification Description
User Interface Color touch screen and keypad operation
Display 7” TFT touch screen 800 x 480 color
Enclosure IP Rating IP31
Power Dedicated lithium battery or power adapter
Battery Rechargeable Li-Ion battery, typically 12 hours of operation, less than 5 hours recharge.
Weight 28 lbs (12.8 kg) without media
Media Sebacate oil
Size 11.81 x 8.66 x 7.56 in (300 x 220 x 192 mm)
Communications USB, LAN, Bluetooth, Wi-Fi and Ethernet
HART Communication Read, congure and calibrate HART devices - DD les updated periodically
Data Storage
8 GB
Data Logging Up to 1,000,000 readings (data and time stamped)
Task Documentation Up to 1,000 tasks can be stored with data
Automation Functions Switch test, auto step, leak test
User Interface Localization English, German, French, Italian, Spanish, Portuguese, Chinese, Japanese, and Russian
Pump Life > 1,000,000 cycles
Environmental Specifications
Operating temperature: 32 °F to 122 °F( 0 °C to 50 °C)
Storage temperature: -20 °C to 60 °C ( -4 °F to 120 °F)
Humidity: <90%, non-condensing
Certification ISO 17025 accredited certicate of calibration with NIST-traceable data
Compliance CE
Software Compatibility ACal, Additel Land and Additel Link for access via mobile application
Warranty 1 year
Hose & Filter End of Life
The estimated End of Life (EOL) expectancy for all hoses and lters (pneumatic and hydraulic) is
approximately 10 years and should be replaced at the rst sign of wear or damage.
Specification
Model
Pressure Range
Accuracy
Pressure Type Media
(psi) (bar)
ADT161-01-GPXX
for 0.01%FS
1,000 70 0.01% FS Gauge G,L
1,500 100 0.01% FS Gauge G,L
2,000 140 0.01% FS Gauge G,L
3,000 200 0.01% FS Gauge G,L
5,000 350 0.01% FS Gauge G,L
10,000 700 0.01% FS Gauge G,L
ADT161-02-GPXX
for 0.02%FS
1,000 70 0.02% FS Gauge G,L
1,500 100 0.02% FS Gauge G,L
2,000 140 0.02% FS Gauge G,L
3,000 200 0.02% FS Gauge G,L
5,000 350
0.02% FS Gauge
G,L
10,000 700 0.02% FS Gauge G,L
05
Corporate Headquarters
2900 Saturn St #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Pressure Calibration Equipment
Phone: 714-998-6899 Email: sales@additel.com
Rev # 20230904
Optional Accessories
Model number Description Picture
ADT107-X-KIT
Includes dual ports manifold, and
zero return communication cable
ADT161
Pressure modules
see pg. 4
ADT100-762-X ADT762 right angle connector
9201 Sebacate oil (1-Liter)
ORDERING INFORMATION
Accuracy:
01 = 0.01% of full span
02 = 0.02% of full span
Pressure port type:
N - 1/4 NPT female
N2 - 1/2 NPT female
B - 1/4 BSP female
B2 - 1/2 BSP female
M - M20X1.5 female
ADT762
01
Model Number
GP10K
N
Accessories (included)
Model number Quantity
Picture
9022
Test Leads
2 sets
(4 pcs)
USB Cable 1 pc
9725
Rechargeable Li-ion battery
1 pc
9818
110V/220V external
Power adapter
1 pc
Funnel 1 pc
ADT100-762Hose
High pressure hose
1 pc
Small accessory case
(For cables and adaptors)
1 pc
Transportation cap 1 pc
9907-762
Carrying case
1 pc
Sebacate oil
(500ml)
1 bottle
O rings for liquid storage tank
3.5*1.2-NBR70
5 pcs
O rings for right angle connector
5*1.5-NBR70
5 pcs
9060
Pressure module connection cable
1 pc
ADT100-762-X
1 pc
ISO17025 accredited
calibration certificate
1 pc