Contamination Prevention System (CPS) Prolongs the
Maintenance Interval of the Controller
Calibration of DUT’s (devices under test) often introduces contaminates to a
calibration system. Contaminates can cause restrictions in valves, lines and
lters. Additel has included a turn-key solution with the ADT793 to help reduce
these concerns and improve durability and dependability when calibrating
customers devices. The use of an automatic contamination prevention system
and integration rmware allows for purges between pressure cycles to further
reduce the possibility of introducing contaminates into the system.
Quick Change Pressure Modules (30 seconds)
Additel's 151 pressure control modules can be installed or replaced within
30 seconds or less. Simply open the upper edge of the cabin to open. As
the door opens, the controller will automatically release pressure, providing
the safe removal and installation of the ADT151 modules. Additel offers ve
different pressure ranges for the ADT793 controller. Select between 1000
bar (15000 psi), 700 bar (10000 psi), 400 bar (6000 psi)
,
350 bar (5000
psi) or 200 bar (3000 psi) with the variety of accuracies to meet application
requirements.
A Singular Pressure Module Design with a Wide Range of
Capabilities
Additel’s 151 pressure modules have been designed with flexibility and
efficiency in mind and can easily be swapped out as needed to meet
changing needs and workloads.Higher pressure ADT151 modules cover
two separate pressure ranges and come with individual calibrations for each
range. This allows for each module to accurately cover a wide range of
pressure workloads. Also, each ADT151 is available in (3) different accuracy
levels (0.02% FS, 0.01% FS and 0.01% of reading) to meet the demanding
needs of our customers.
External Reservoir Support
High volume high pressure calibration work can quickly consume working
media which can increase hands-on time and can be a hassle for laboratory
personnel. The ADT793 supports switching between internal and external
liquid storage tanks, allowing users to quickly connect the large capacity
external liquid reservoir saving time and labor.
20% Pressure Step within 30 Seconds
In the process of efcient and fast-paced production line testing, verication
and calibration, users have strict requirements on the speed of pressure
controllers. ADT793 adopts professional control technology to effectively
improve control rate and stability: control response time (typical) ≤30
seconds, control stability (typical) ≤±(0.003~0.005)%FS
,
see specications
for more details.
Load Volumn Expander Expands the Loading Volume to 250ml
ADT793 supports reciprocating pressurization through a load volume
expander to expand the loading volume to 250ml, which can meet the needs
of production line users to test multiple UUTs at the same time.