Thermocouple Calibration Furnaces
OVERVIEW
Temperature control from 100°C to 1210°C
Two models to choose from:
Reference (ADT878) and Standard (ADT875)
Display Accuracy of ±1.5°C (ADT878)
Stability of ±0.1°C
4 on-board measurement channels (PC option)
Process calibrator option provides a multi-channel
readout for TCs, switches and transmitters, including
task documentation and HART communication
Portable, rugged and quick to temperature
Self-calibration feature (PC option)
Multi-zone temperature control
Internal and external sensor control (PC option)
Metallic interchangeable inserts
Wi-Fi and Bluetooth capable
Color touch screen display
ISO 17025-accredited calibration w/data included
Patent pending technology
We understand the many challenges associated with thermocouple calibration work. That is precisely
why we decided to introduce the ADT875-1210 and ADT878-1210 Thermocouple Calibration
Furnaces.
With an unmatched stability, uniformity and an optional on-board process calibrator, calibrating
thermocouples has never been easier. With two separate units to choose from, the ADT875-1210 and
ADT878-1210 furnaces include a patented multi-zone temperature control which provides a never
before seen, highly stable and uniform heat source to ensure you get the best possible results from a
modest investment. With metallic interchangeable inserts, users have the exibility needed to service
a wide variety of UUT’s and the durability they have come to expect from Additel. The ADT875-1210
and ADT878-1210 can be purchased with or without our on-board process calibration electronics to
provide exibility for customers who are needing the best 1200°C heat source on the market.
If thermocouple calibration and/or verication work is part of your workload, you don’t want to miss out
on this opportunity to save valuable time and money with these best in class furnaces from Additel.
ADT875 and ADT878
02
Corporate Headquarters
2900 Saturn Street #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Temperature Calibration Equipment
Phone: 714-998-6899 Fax: 714-998-6999 Email: sales@additel.com
Rev # 20230407
Temperature Control
General Specifications
Specification 875-1210 878-1210
[1]
Temperature
Range
100°C to 1210°C
Display
Accuracy
±1.2°C @ 100°C
±1.2°C @ 300°C
±1.2°C @ 600°C
±1.6°C @ 900°C
±2.0°C @ 1210°C
±1.0°C @ 100°C
±1.0°C @ 300°C
±1.0°C @ 600°C
±1.2°C @ 900°C
±1.5°C @ 1210°C
Stability ±0.1°C
Axial Uniformity
(20mm zone)
±0.6°C @ 100°C
±1.2°C @ 300°C
±1.5°C @ 600°C
±1.5°C @ 900°C
±1.5°C @ 1210°C
±0.4°C @ 100°C
±0.8°C @ 300°C
±1°C @ 600°C
±1°C @ 900°C
±1°C @ 1210°C
Radial
Uniformity
±0.2°C @ 100°C
±0.3°C @ 300°C
±0.4°C @ 600°C
±0.8°C @ 900°C
±1°C @ 1210°C
±0.2°C @ 100°C
±0.3°C @ 300°C
±0.4°C @ 600°C
±0.6C @ 900°C
±0.8°C @ 1210°C
Loading Effect ±0.5°C
Environmental
Conditions
8°C to 38°C guaranteed accuracy
0°C to 50°C, 0% to 90% RH non-condensing,
3000 M altitude for normal operation
Storage
Conditions
-20°C to 60°C
Immersion Depth
XR style inserts = 138 mm (5.43")
XS style inserts = 116 mm (4.57")
(see insert ordering info for more details)
Insert Size - OD 24.8 mm (0.98 inches)
Heating Time 50 min: 23°C to 1210°C
Cooling Time
50 mins:1210°C to 300°C
50 mins: 300°C to 50°C
55 mins:1210°C to 300°C
55 mins: 300°C to 50°C
Typical Time to
Stability
15 min
Resolution 0.01°C
Units °C, °F, and K
Display 6.5 in (165 mm) color touch screen
Size (H x W x D) 345 x 170 x 330 mm (13.6 x 6.7 x 13.0 in)
Weight 10.6 kg (23.4 lbs)
The Additel ADT875 & ADT878 Thermocouple Calibration
Furnaces have been designed with a unique and innovative
way of controlling temperature and temperature gradients.
We like to call it “Advanced Adaptive Control”. This exciting
new design feature incorporates our patent pending wind
tunnel control technology with Additel’s impressive 3-zone
temperature control to provide the very best uniformity and
stability possible.
Each ADT875 & ADT878 is tested and calibrated in Additel’s
accredited laboratory (Brea, CA) to ensure that each unit
is ready to go when the customer opens the package. The
included accredited calibration certificate provides data
relating to accuracy, stability and uniformity to help provide
even more confidence in the testing and calibration of each
and every ADT875 & ADT878 Thermocouple Calibration
Furnace.
Specification 875-1210
878-1210
Power Requirements 90-254 VAC, 45-65 Hz, 580 W
Mechanical Testing
Vibration: 2 g (10-500 Hz), 30 min for 2 sides
Impact: 4 g three times
Drop test: 500 mm (19.6 in)
Communication USB A, USB B, RJ45, WiFi, Bluetooth
Localization English, Chinese, Japanese, Russian, German
Warranty 1 year
[1] 878-1210 specications require the use of an "XR" style insert.
Otherwise default to the 875-1210 specications.
Temperature Calibration Equipment
03
Corporate Headquarters
2900 Saturn Street #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Phone: 714-998-6899 Fax: 714-998-6999 Email: sales@additel.com
Rev # 20230407
Process Electronics
Both the ADT875 & ADT878 can be ordered with Additel's Process Calibrator
(PC) option. The Process Calibrator Option combines the many features found
in a thermocouple readout device and process calibrator with the ADT875 &
ADT878 Calibration Furnaces.
This unique option includes Additel’s patented Quick-Push connectors which
accommodate virtually all TC connection types. The process option also
includes the ability to measure a reference grade thermocouple and up to
(4) under test channels. Channels 1 and 2 can measure mA, voltage, perform
switch testing and source 24V DC. In addition to these measurement functions,
the process option provides full documenting capability of creating tasks, saving
"as found" and "as left" results and HART communications for simplied
transmitter work. The snapshot feature allows users to capture all information
displayed on the screen with a touch of the screen. This optional add-on allows
for data logging of all channels using our auto step and a ramp functions.
By utilizing the external reference option users can select to control to the
furnace set point using an external control probe, which helps to reduce
uncertainties. The external control probe feature also facilitates the
handy self-calibration feature!
Input Specifications (Process Calibrator [PC] Option)
TC Measurement Specifications and Calculations
(Process Calibrator [PC] Option)
TC Type
TEMP
(
°
C)
Error (
°
C)
[1]
TC Type
TEMP
(
°
C)
Error (
°
C)
[1]
875 878 875 878
K
(CH1-
CH4)
100 ±0.182 ±0.172
S
(CH1-
CH4)
100 ±1.102 ±1.094
300 ±0.266 ±0.236 300 ±0.924 ±0.899
600 ±0.310 ±0.251 600 ±0.888 ±0.837
900 ±0.397 ±0.304 900 ±0.868 ±0.793
1210 ±0.517 ±0.382 1210 ±0.865 ±0.765
N
(CH1-
CH4)
100 ±0.273 ±0.264
R
(CH1-
CH4)
100 ±1.080 ±1.072
300 ±0.270 ±0.243 300 ±0.869 ±0.844
600 ±0.309 ±0.256 600 ±0.804 ±0.755
900 ±0.368 ±0.285 900 ±0.771 ±0.699
1210 ±0.455 ±0.335 1210 ±0.766 ±0.670
E
(CH1-
CH4)
100 ±0.136 ±0.126
B
(CH1-
CH4)
250 ±3.182 ±3.170
300 ±0.153 ±0.130 300 ±2.645 ±2.631
600 ±0.210 ±0.154 600 ±1.409 ±1.379
900 ±0.291 ±0.202 900 ±1.049 ±1.003
1000 ±0.297 ±0.196 1210 ±0.905 ±0.839
L
(CH1-
CH4)
100 ±0.223 ±0.214
T
(CH1-
CH4)
100 ±0.194 ±0.185
300 ±0.271 ±0.241 300 ±0.191 ±0.166
600 ±0.308 ±0.251 400 ±0.217 ±0.183
900 ±0.522 ±0.448
S
(EXT.
REF)
100 ±0.277 ±0.273
U
(CH1-
CH4)
100 ±0.270 ±0.261 300 ±0.242 ±0.229
300 ±0.189 ±0.164 600 ±0.249 ±0.224
600 ±0.227 ±0.176 900 ±0.258 ±0.220
J
(CH1-
CH4)
100 ±0.186 ±0.177 1210 ±0.266 ±0.216
300 ±0.197 ±0.168
R
(EXT.
REF)
100 ±0.271 ±0.266
600 ±0.256 ±0.200 300 ±0.228 ±0.216
900 ±0.281 ±0.197 600 ±0.227 ±0.202
1200 ±0.414 ±0.294 900 ±0.230 ±0.194
1210 ±0.240 ±0.192
[1] Excluding cold junction compensation errors.
ADT875 & ADT878 Process Calibrator [PC] option electronics
Specification 875-1210 878-1210
TC Measurement Channels
Patented TC terminals:
Accepting S, R, K, B, N, E, J, T, L, and U
TC Measurement Accuracy
Type K
Ch. 1-4 (excluding sensor)
±0.182°C @ 100°C
±0.266°C @ 300°C
±0.310°C @ 600°C
±0.397°C @ 900°C
±0.517°C @1210°C
±0.172°C @ 100°C
±0.236°C @ 300°C
±0.251°C @ 600°C
±0.304°C @ 900°C
±0.382°C @ 1210°C
TC Range
–75 mV to 75 mV (UUT Channels 1-4)
–18 mV to 18 mV (Reference Channel)
TC Resolution 0.0001 mV, Input Impedance < 10Ω
TC Voltage Accuracy
0.02% RD + 8μV (ch. 1-4)
0.01% RD + 2μV (ref ch.)
0.01% RD + 8μV (ch. 1-4)
0.005% RD + 2μV (Ref ch.)
Internal CJC Accuracy
±0.35°C (ch. 1-4)
±0.25°C (ref ch.)
±0.30°C (ch. 1-4)
±0.20°C (ref ch.)
Current Range
–30 mA to 30 mA
Current Accuracy ±(0.02% of rdg+ 2μA)
±(0.01% of rdg + 2μA)
Current Resolution 0.0001 mA
Voltage Ranges –12 V to 12 V and –30 V to 30 V
Voltage Accuracy ±(0.02% of rdg+ 2mV) ±(0.01% of rdg+ 0.6mV)
Voltage Resolution 0.0001 V, Input Impedance >1MΩ
DC 24V Output 24 V ± 10%, MAX 60 mA
Hart Communication Optional (ADT875PC and ADT878PC Models)
Temperature Coefficient
0°C to 8°C and 38°C to
50°C
TC Readouts: ±5 ppm/°C
Current: ±5 ppm/°C
Voltage: ±5 ppm/°C
Switch Test Mechanical or Electrical - Channels 1 & 2 only
Documentation
Up to 1,000 tasks which store up to 10 results each
containing as found and as left data. Snap shot feature
allows for screen captures. Records auto step and ramp
functions
04
Corporate Headquarters
2900 Saturn Street #B
Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101
Pleasant Grove, UT 84062, USA
Temperature Calibration Equipment
Phone: 714-998-6899 Fax: 714-998-6999 Email: sales@additel.com
Rev # 20230407
Optional Accessories
Model Description Picture
9915-878
Carry case for ADT875-1210
or ADT878-1210
with wheels
ADT110-87X-TC-
INSERT-XX
Insert for ADT875-1210 or
ADT878-1210
(see insert ordering
information below)
AM1210-12
Reference TC - Type S:
Platinum/10% Rhodium vs.
platinum - 12" length
(see AM1210 specs below)
9080
Cable Kit (includes TC
plug, compensation cable,
S,R,K,J,T,E,N)
AM1210-12 Type S Reference Standard Thermocouple
Temperature Range 0°C to 1300°C
Type
Type S: Platinum/10% Rhodium vs.
platinum
Long Term Drift
±0.5°C at 1084.62°C after 1 year
typical use
Short Term stability ±0.2°C at 1084.62°C
Diameter of thermocouple wire 0.5 mm
Sheath Material Alumina
Sheath Dimensions
OD: 6 mm (0.236");
Length: 305 mm (12.0")
Protective Carrying Case Included
Documentation Report of test with data
Insert Information
[1] Insert models ending in the letter S have probe holes of shallower depths. Please call with questions.
Note: ISO 17025 accredited probe calibration available, contact Additel for
more information"
Ordering Information
Model:
ADT875, ADT875PC
ADT878, ADT878PC
ADT875
Model Number
Temperature Range (at 23
):
1210: 100
°C to 1210°C
AR
110 V
Insert Model:
AR, BS, CS, DR,
ES, FS, GS, HR
Power Option:
110 V
220 V
1210
Standard Accessories
Model Quantity Picture
Calibration Furnace
and selected Insert &
insulator
1 pc.
Power cable 1 pc.
USB Cable 1 pc.
Insert removal tool 1 pc.
Test leads
(PC option only)
2 sets (6 pcs.)
Accredited Calibration
Certication
1 pc.
Manual 1 pc.
Accessories
ADT110
Insert Ordering Information
875 TC
INSERT
Insert Model:
AR, BS, CS, DR,
ES, FS, GS, HR
AR
Insert Type:
ADT875
ADT878
* ADT878-1210 specications require the use of an "XR" style insert
* ADT878-1210 specications require the use of an "XR" style insert
Reference Style inserts - 138 mm (5.43'') hole depth
- for use with both 878 and 875 models
Short Style Insert - 116 mm (4.57'') hole depth - Only for the ADT875 - 1210
Model Specification Model Specification Model Specification Model Specification Model Specification
AR HR CS GS BS
DR FS ES
[1]
[1]
[1]
[1]
[1]