AccuMac AM1760 & AM1762
Secondary SPRTs
OVERVIEW
The AM1760 series Secondary SPRTs provides our customers with reliable secondary standards that can be used
daily in their labs. These SPRTs feature an accuracy of <0.006°C at 0.01°C, a short term stability of <0.002°C and a
very low drift rate of less than 0.004°C after 100 hours at 661°C. Two different lengths of these SPRTs are available at
either12 inches (305 mm) or 20 inches (508 mm).
The sensing element is designed to protect the platinum sensing wire from contamination at high temperatures, giving
these SPRTs a high level of stability and repeatability in performance. A uniquely designed support structure and lling
material provides excellent balance between the hysteresis effect, mechanical shock and thermal shock performance.
This high performance probe fully meets the ITS-90 criteria for reference thermometers. Each probe comes standard
with an ISO 17025 accredited calibration.
SPECIFICATIONS
Reliable SPRT for Daily Use
Rtpw Drift: <4mK after 100hrs at 661
°
C
Short-term Stability : <0.002
°
C at 0.01
°
C
Temperature Range: -200
°
C to 670
°
C
Inconel™ Sheath to Withstand Harsh Environments
Fully Meets the ITS-90 Criteria for Reference Thermometers
ISO 17025 Accredited Calibration Included
AM1760 AM1762 AM1760
/
AM1762-BEND
Temperature Range -200°C to 670°C
Nominal Resistance at 0.01°C 100 Ω 25 Ω 100 Ω / 25 Ω
Temperature Coefcient 0.003925 Ω/Ω/°C
Calibrated Accuracy (k=2)
[1][2]
±0.010°C at -196°C
±0.006°C at 0.01°C
±0.015°C at 420°C
±0.025°C at 661°C
Long Term Drift at 0.01°C
[3]
<0.004°C at TPW after 100 hours at 661°C
Short Term Stability <0.002°C
Thermal Shock <0.002°C after 10 thermal cycles from minimum to maximum temperatures
Self-heating 0.0015°C at 0.5 mA current 0.0015°C at 1 mA
0.0015°C at 0.5 mA
0.0015°C at 1 mA
Response Time
[4]
9 seconds
Measurement Current 0.5 mA 1 mA 0.5 mA / 1 mA
Sensor Length 1.65 in (42 mm)
Sensor Location 0.2 in (5 mm) from tip
Insulation Resistance >1000 MΩ at room temperature
Sheath Material Inconel™
Sheath Dimensions
1760-20: 0.25 in (6.35 mm) (OD)
x 20 in (508 mm) (L)
1760-12: 0.25 in (6.35 mm) (OD)
x 12 in (305 mm) (L)
1762-20: 0.25 in (6.35 mm) (OD)
x 20 in (508 mm) (L)
1762-12: 0.25 in (6.35 mm) (OD)
x 12 in (305 mm) (L)
0.25 in (6.35 mm) (OD)
x 9.75 in (247.65 mm) vertical
x 4.5 in (114.3 mm) horizontal
External Leads Teon™ insulated copper wire, 4 leads, 3.5 feet (2 meters)
Handle Dimension 0.59” (15 mm) (OD) x 2.56” (65 mm) (L)
Handle Temperature Range -50°C to 180°C
Calibration ISO 17025 accredited calibration
[1] Includes calibration and 100 hour drift.
[2] Probe calibration ranges may differ from probe temperature ranges (see Calibrated Accuracy for calibration ranges).
[3] Long-term drift rate is for reference only. It could be affected by such facts as handling, application, and maintenance, etc.
[4] For 63% response to step change in water moving at 1 meter per second.